Evaluation of the Potentiometric Titration Used to determinate arsenic)III( in mixed and aqueous media comparing with Atomic Absorption Method

تقييم طريقة المعايرة الكمونية لتعيين تركيز الزرنيخ (III) في الأوساط المائية والمختلطة بالمقارنة مع طريقة مطيافية الامتصاص الذري

2189   0   54   0 ( 0 )
 Publication date 2014
and research's language is العربية
 Created by Shamra Editor




Ask ChatGPT about the research

This study aims to compare the results of two methods ,used to determinate arsenic in mixed and aqueous media: proposed potintiometrc titration and referenced atomic absorption and then applied the proposed method to some environmental samples. In this paper, Arsenic)III( has been titrated with iodine in aqueous and mixed media using the direct method and platinum electrode after choosing optimal technical and analytical conditions. The atomic absorption, which is often used to determine the arsenic in many applications, has been chosen as comparison guide, where the reaction take place between sodium borate (NaBH4) and arsenic Acid H3AsO3 to produce the effective compound AsH3 used to determine arsenic by this instrument . Different statistical treatments were used to evaluate the result of the determination of arsenic in desired media. The result of this investigation suggest that , it’s highly recommend to use the proposed method for its high efficiency, ease, simplicity and low cost comparing with atomic absorption technique that requires expensive equipment


Artificial intelligence review:
Research summary
تهدف هذه الدراسة إلى مقارنة طريقتين لتحديد تركيز الزرنيخ (III) في الأوساط المائية والمختلطة: طريقة المعايرة الكمونية وطريقة مطيافية الامتصاص الذري. تم استخدام مسرى البلاتين لمعايرة الزرنيخ مع اليود في الأوساط المائية والمختلطة بعد تحديد الشروط التحليلية والتقنية المثلى. كما استخدمت تقنية مولد الهيدرات في مطيافية الامتصاص الذري، حيث يتفاعل بورات الصوديوم مع حمض الزرنيخ لإنتاج مركب AsH3 الفعال. أظهرت النتائج أن طريقة المعايرة الكمونية تتميز بكفاءة عالية وسهولة في الاستخدام وتكلفة منخفضة مقارنة بمطيافية الامتصاص الذري التي تتطلب تجهيزات غالية. تم تطبيق الطريقة المقترحة على عينات بيئية مختلفة وأظهرت نتائج موثوقة ودقيقة. توصي الدراسة باستخدام طريقة المعايرة الكمونية كبديل فعال لمطيافية الامتصاص الذري في تحديد تركيز الزرنيخ في الأوساط المختلفة.
Critical review
تعتبر هذه الدراسة مفيدة ومهمة في مجال تحليل الزرنيخ، حيث تقدم طريقة بديلة وفعالة من حيث التكلفة والسهولة مقارنة بمطيافية الامتصاص الذري. ومع ذلك، يمكن تقديم بعض الملاحظات النقدية لتحسين البحث. أولاً، قد يكون من المفيد توسيع نطاق الدراسة لتشمل أنواع أخرى من الأوساط البيئية والمختلطة. ثانياً، يمكن تعزيز الدراسة بإجراء تجارب إضافية للتحقق من موثوقية الطريقة المقترحة في ظروف مختلفة. أخيراً، قد يكون من المفيد إجراء مقارنة مع طرق تحليلية أخرى مثل التحليل بالتنشيط النيوتروني أو تقنيات البلازما المقرونة حثياً لتقديم صورة شاملة عن فعالية الطريقة المقترحة.
Questions related to the research
  1. ما هي الطريقتان المستخدمتان في هذه الدراسة لتحديد تركيز الزرنيخ (III)؟

    الطريقتان المستخدمتان هما المعايرة الكمونية ومطيافية الامتصاص الذري.

  2. ما هي المزايا التي تتمتع بها طريقة المعايرة الكمونية مقارنة بمطيافية الامتصاص الذري؟

    طريقة المعايرة الكمونية تتميز بكفاءة عالية وسهولة في الاستخدام وتكلفة منخفضة مقارنة بمطيافية الامتصاص الذري التي تتطلب تجهيزات غالية.

  3. ما هي الأوساط التي تم تطبيق الطريقة المقترحة عليها في الدراسة؟

    تم تطبيق الطريقة المقترحة على عينات بيئية مختلفة مثل مياه الصرف الصحي والتربة والنباتات.

  4. ما هي التوصية الرئيسية التي توصلت إليها الدراسة؟

    توصي الدراسة باستخدام طريقة المعايرة الكمونية كبديل فعال لمطيافية الامتصاص الذري في تحديد تركيز الزرنيخ في الأوساط المختلفة.


References used
Luong,H .; Majid, E .; Male, K.Analytical Tools for Monitoring Arsenic in the Environment.Open Analytical Chemistry Journal, 1, 2009, 7-14
Nurulhaidah, D.; Nor , Y.; Tan ,T.; Abdul , A.Electrochemical Sensor for As(III) Utilizing CNTs/ Leucine/Nafion Modified Electrode.Int. J. Electrochem. Sci., 7 ,2012, 175 –185
Kazimierz ,W.; Katarzyna, W.; Bryan, P.; Sasi,S.; Joseph, A .Determination of As(III), As(V), monomethylarsonic acid,dimethylarsinic acid and arsenobetaine by HPLC_ICP_MS: analysis of reference materials, fish tissues and urine. Talanta 58 (2002) 899_907
Niedzielski, P.; Siepak, M.; Siepak, J.; Przybyłek, J.; Determination of Different Forms of Arsenic, Antimony and Selenium in Water Samples Using Hydride Generation. Polish Journal of Environmental Studies Vol. 11, No. 3 (2002), 219-224
Pereira1, S.; Ferreira, L.; Oliveira, G.; Palheta, D.; Barros, B. Spectrophotometric determination of arsenic in soil samples using 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (Br- PADAP). Ecl. Quím., São Paulo, 33(3): 23-28, 2008
rate research

Read More

This research aims to study the effect of certain analytical conditions to determine the arsenic in aqueous media using a Potentiometric, and then choose and adopt the optimum conditions for this determinism. We studied the effect of each of the (PH value, type, concentration puffer, effect of crippling electrolytes) to determine the arsenic in aqueous media and calibration using a standard solution of iodine. We also worked on raising the sensitivity of the method detection limit access to the ppm. It should be noted that technical conditions must be tuned to determine the actual value of the end point calibration and data processing of standard solution (initial velocity and deceleration near the end point calibration). The study showed that the PH value appropriate for calibration of arsenic with iodine (8-9), also found that the response process of arsenic on the electrode platinum has been excellent when using Na2SO4 as puffer with concentration of 0.1 M, which has been successfully used to quantify arsenic As+3 up to detect up to 2 × 10-6M. As it turns out during the study, it was not possible to determine the arsenic presence of copper ions in the solution, while the presence of lead and sulfate ions and chloride was not affective.
This research aims to study the effect of certain analytical conditions to determine the arsenic in in mixed media (water _ alcohol ) where the user either alcohol ( methanol, ethanol , propanol using a Potentiometric, and then selected as conditi ons optimal certified for this determine. The effect of each of the (PH value, type, concentration electrolyte, effect of crippling electrolytes) to determine the arsenic in mixed media ( 50:50 water _ alcohol ) and calibration using a standard solution of iodine.In addition to that, there are technical conditions must be tuned to determine the actual value of the end point calibration and data processing of standard solution (initial velocity and deceleration near the end point calibration). The study showed that the PH value appropriate for calibration of arsenic with iodine (pH=8), also found that the response process of arsenic on the electrode platinum has been excellent when using NaHCO3 as electrolyte with concentration of 1 M, which has been successfully used to quantify arsenic As+3, where the value reached detection limit of arsenic to (1×10-5M) in media (50:50water _ methanol ) , but when you use a media ( 50:50 water _ ethanol ) arrived to the limit of detection (1×10-3 M ), As for propanol arrived to the limit of detection (1×10-2 M). Was calculated as the standard deviation and standard deviation percentile confidence limit in addition to the feedback that indicates the efficiency of the method of determining the concentration of arsenic and detection limit for each medium.
This research study the effect of the proportion of the media (water - alcohol ) to determine where the arsenic used alcohol ( methanol _ ethanol _ propanol ) Using a Potentiometric method. It conducted the calibration process for the specific size o f the solution studied a specific size of the standard solution of iodine Using a Potentiometric method , after the application of analytical conditions and technical optimal to determine arsenic , and then we studied the effect of mixing ratio on the detection limit for arsenic in mixed media (water _ Alcohol ) . The study showed that the value of the detection limit of arsenic vary with the type of alcohol used and accounted for in the reaction medium , it reached to (1× 10-7 M) in media (75:25 water _ methanol ) , but when you use a media ( 75:25 water _ ethanol ) arrived to the limit of detection (1×10-4 M ), As for propanol arrived to the limit of detection (1×10-3 M). where the value reached detection limit of arsenic to (1× 10-5 M) in media ( 50:50 water _ methanol ) , but when you use a media ( 50:50 water _ ethanol ) arrived to the limit of detection (1×10-3 M ), As for propanol arrived to the limit of detection (1×10-2 M). The study showed that the use of the ratio ( 75:25 water _ alcohol ) gave a higher value limit of detection and therefore better results . Was calculated as the standard deviation and standard deviation percentile confidence limit in addition to the feedback that indicates the efficiency of the method of determining the concentration of arsenic detection limit for each medium.
The aim of this study was to determine the trace of some heavy metals elements pb, Cd, Cu, Zn & Fe in the leaves and periderm of the Avundodonax L. using Atomic absorption spectroscopy. The samples were collected from twenty locations and divided into four groups according to traffic and human activities. The results showed an excess in the concentration of lead, cadmium and copper in the leaves and periderm of the Avundodonax L., which enables the use of this plant as a vital indicator for the pollution by these heavy metals. The concentration of cadmium in the leaves and periderm ranged between (0.231- 6.278ppm) and for lead between(1.057- 47.592ppm) and copper between(1.022 -45.771ppm). While the concentration of both zinc and iron remained within the permissible natural limits.
The objective of this study is to study some of the optimal analytical conditions for the determination of the Cu2+ ion using potentiometric titration method using the platinum electrode as a working electrode and the AgCL/Ag as a Reference electro de. In this study, the technical conditions were studied (working Electrode type, Reference Electrode type, Direction of titration, delay time, Proportional band, Speed of stirring) and the analytical conditions (pH value, time factor, quantity of solid KI added, electrolyte type and its concentration, detection limit) were studied using oxidation - reduction reactions as the method of iodide titration which applied successfully electrochemical for determine the Cu2+ ion.
comments
Fetching comments Fetching comments
Sign in to be able to follow your search criteria
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا