Study of the Effect of Electrostatic Charge on the Delivered Dose of Metered Dose Inhaler

دراسة تأثير الكهرباء الساكنة على الجرعة المتحررة من شكل إرذاذي معاير الجرعة

1786   0   14   0 ( 0 )
 Publication date 2013
and research's language is العربية
 Created by Shamra Editor




Ask ChatGPT about the research

A study of the effect of electrostatic charge that is generated in actuator due to drying methods on the delivered dose of metered dose inhaler has been done in tow cases: Washing actuator and drying it by a clothe. Washing actuator and let it dry alone. A study has been done on metered dose inhaler containing salbutmaol sulfate and manufactured by a local company. The results show the importance of drying methods on the delivered dose, the delivered doses of metered dose inhaler after washing the actuator and dry it by a clothe hadn't been achieved the USP-34 requirements (out of range 75-125%), while the delivered doses after washing actuator and let it dry alone had been achieved the USP-34 requirements. The difference in results between the two cases belong to electrostatic charge generated in actuator due to drying methods


Artificial intelligence review:
Research summary
تتناول الدراسة تأثير الكهرباء الساكنة المتولدة في الحاضنة (actuator) على محتوى الجرعة المتحررة من عبوة شكل إرذاذي معاير الجرعة (metered dose inhaler) التي تحتوي على كبريتات السالبوتامول. تم إجراء التجارب في حالتين: الأولى بعد غسل الحاضنة وتجفيفها باستخدام قطعة قماش، والثانية بعد غسل الحاضنة وتركها تجف في الهواء. أظهرت النتائج أن الطريقة المتبعة في تجفيف الحاضنة تؤثر بشكل كبير على محتوى الجرعة المتحررة. في الحالة الأولى، لم تحقق الجرعات المتحررة المتطلبات الدستورية وكانت خارج المجال المقبول (75-125%)، بينما في الحالة الثانية، حققت الجرعات المتحررة المتطلبات الدستورية وكانت ضمن المجال المقبول. يعزى هذا الاختلاف إلى الكهرباء الساكنة المتولدة في الحاضنة جراء عملية التجفيف باستخدام قطعة قماش. توصي الدراسة بترك الحاضنة تجف تلقائياً للحصول على جرعات ثابتة ومطابقة للمعايير الدستورية.
Critical review
دراسة نقدية: تعتبر هذه الدراسة مهمة لأنها تسلط الضوء على تأثير الكهرباء الساكنة على الجرعات المتحررة من أجهزة الاستنشاق، وهو موضوع لم يتم دراسته بشكل كافٍ في الأدبيات السابقة. ومع ذلك، يمكن توجيه بعض النقد البنّاء للدراسة. أولاً، كان من الممكن أن تكون الدراسة أكثر شمولاً إذا تم تضمين أنواع أخرى من الأدوية أو أجهزة الاستنشاق. ثانياً، لم يتم التطرق إلى تأثير العوامل البيئية الأخرى مثل الرطوبة ودرجة الحرارة على توليد الكهرباء الساكنة. أخيراً، كان من الممكن أن تكون الدراسة أكثر قوة إذا تم إجراء تجارب إضافية للتحقق من النتائج في ظروف مختلفة. على الرغم من هذه النقاط، فإن الدراسة تقدم إسهاماً قيماً في فهم تأثير الكهرباء الساكنة على الجرعات المتحررة من أجهزة الاستنشاق.
Questions related to the research
  1. ما هو الهدف الرئيسي من الدراسة؟

    الهدف الرئيسي من الدراسة هو تحديد تأثير الكهرباء الساكنة المتولدة في الحاضنة على محتوى الجرعة المتحررة من عبوة شكل إرذاذي معاير الجرعة.

  2. ما هي الطريقة التي أظهرت نتائج أفضل في تحقيق المتطلبات الدستورية؟

    الطريقة التي أظهرت نتائج أفضل هي غسل الحاضنة وتركها تجف تلقائياً في الهواء، حيث حققت الجرعات المتحررة المتطلبات الدستورية وكانت ضمن المجال المقبول.

  3. ما هو السبب الرئيسي لاختلاف محتوى الجرعات بين الحالتين المدروستين؟

    السبب الرئيسي لاختلاف محتوى الجرعات بين الحالتين هو الكهرباء الساكنة المتولدة في الحاضنة جراء عملية التجفيف باستخدام قطعة قماش.

  4. ما هي التوصية الرئيسية التي قدمتها الدراسة للمستخدمين؟

    التوصية الرئيسية التي قدمتها الدراسة هي ترك الحاضنة تجف تلقائياً للحصول على جرعات ثابتة ومطابقة للمعايير الدستورية.


References used
Anhoj, J.;Bisgaard, H.;Lipworth.B.J. Effect of Electrostatic Charge in Plastic Spacers on the Lung Delivery of HFA-Salbutamol in Children. Br J Pharmacol.47(3),1999,333-336
Anouk, C.A.; Sennbauser, F.H.; Wildhaber, J.H. Factor Affecting the Efficiency of Aerosol Therapy with Pressurized Metered Dose Inhalers through Plastic Spacers. SwissMed Wlky,2001,14-18
SwissMed Wlky,2001,14-18. 3. Deveswaran, R.; Bharath, S.; Basavaraj, BV.; Abraham, S.; Furtado, SH.; Madhavan, V. Concept and Techniques of Pharmaceutical Powder Mixing Process: A Current Update. Research J.Pharm and Tech.2(2).April,2009,245-249
Kwok, P.C.; Chan, H.H. Electrostatic of Pharmaceutical Aerosols. J Pharma Pharmacol.Dec;61(12),2009,1587-99.Doi:10.1211/jpp/61.12.0002
Mitchell, J.P.; Coppolo, D.P.; Nagel, M.W. Electrostatics and Inhaled Medications: Influence on Delivery via Pressurized Metered Dose Inhalers and Add on Devices. Respir Care,2007,Mar;52(3),283-300
rate research

Read More

To calculate the dose distributions of 6 MeV photon beam at variable depth in 3D water phantom the MCNP4C2 code was used, and the simulated dose profile was compared with that of the treatment planning computer system (TPS), and a good agreement w as found between them. In conclusion, the MCNP4C2 code package presents a good tool adaptable to get dose distributions for the 6MeV photon beam and it can be considered as confirmed method for patient dose calculations.
In this article we studied two types of teeth, spherical and conical one. We compared the results with a previous study for the flat tooth. the study of stresses field under these three teeth show that in the soft rocks we prefer to use conical to oth because it is easy to be Implanted in the rock, and this was confirmed by the values of stresses under the tooth. Either in hard rock and due to our need for big values of pressure to shatter the rock we prefer to use spherical tooth because of its property of forming tow cons of stresses under tooth and therefore can be good in drilling hard rock..
The measurements carried out in many medical centers in the middle region) Homs and Hama ( to estimate medical radiation dose which patient received it resulting from some radiation radiographic devices.
In this research a study effect of absorbed surface shape for solar air collector (Flat V-Corrugated ,Sinusoidal Wave-Corrugated ,Rectangular-Corrugated) titled to the horizontal by an angle(30) on forced heat transfer and skin friction coefficient by Fluent Numerical Program at constant solar heat flax (530W/m2) with Reynold’s number (5000≤Re≤8000). The government equation (mass, momentum and energy) are using Finite Volume by (Fluent 6.3) software for considering steady state, two dimensional and turbulent flow. The results show that the different between inlet and outlet air temperature is(13.65oC) and best heat transfer coefficient enhancement is(63%) for (V-Corrugated plate) compared with the flat plate at (Re=8000). Also the results show that the maximum skin friction coefficient is occurred at (V-Corrugated) Cf=0.22 at (Re=8000).
This paper presents an experimental investigation of the flow pattern and wall pressure due to the change of the opening discharge diameter in cylindrical silos. The experiments were performed on a metal silo prototype using corn as a granular mater ial. The silo prototype used in the study was built of a metallic cylinder, which can be opened longitudinally allowing the observation of the flow pattern through a plexi-glass wall, and the wall pressure distribution was measured using a special granular pressure transducer. The present study revealed unexpected results to what might be expected through current theoretical knowledge.
comments
Fetching comments Fetching comments
Sign in to be able to follow your search criteria
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا