High-precision efficiency calibration of a high-purity co-axial germanium detector

152   0   0.0 ( 0 )
 Added by Bertram Blank
 Publication date 2014
  fields Physics
and research's language is English




Ask ChatGPT about the research

A high-purity co-axial germanium detector has been calibrated in efficiency to a precision of about 0.15% over a wide energy range. High-precision scans of the detector crystal and gamma-ray source measurements have been compared to Monte-Carlo simulations to adjust the dimensions of a detector model. For this purpose, standard calibration sources and short-lived on-line sources have been used. The resulting efficiency calibration reaches the precision needed e.g. for branching ratio measurements of super-allowed beta decays for tests of the weak-interaction standard model.



rate research

Read More

229 - B. Blank , P. Ascher , M. Gerbaux 2020
Following work done in the energy region above 100 keV, the high-precision calibration of a co-axial high-purity germanium detector has been continued in the energy region below 100 keV. Previous measurements or Monte-Carlo simulations have been repeated with higher statistics and new source measurements have been added. A precision as in the high-energy part, i.e. an absolute precision for the detection efficiency of 0.2%, has been reached. The low-energy behaviour of the germanium detector was further scrutinized by studying the germanium X-ray escape probability for the detection of low-energy photons. In addition, one experimental point, a gamma ray at 2168 keV from the decay of 38K, has been included for the total-to-peak ratios agreeing well with simulations. The same gamma ray was also added for the single- and double-escape probabilities. Finally, the long term stability of the efficiency of the germanium detector was investigated by regularly measuring the full-energy peak efficiency with a precisely calibrated 60Co source and found to be perfectly stable over a period of 10 years.
Large, high-purity, germanium (HPGe) detectors are needed for neutrinoless double-beta decay and dark matter experiments. Currently, large (> 4 inches in diameter) HPGe crystals can be grown at the University of South Dakota (USD). We verify that the quality of the grown crystals is sufficient for use in large detectors by fabricating and characterizing smaller HPGe detectors made from those crystals. We report the results from eight detectors fabricated over six months using crystals grown at USD. Amorphous germanium (a-Ge) contacts are used for blocking both electrons and holes. Two types of geometry were used to fabricate HPGe detectors. As a result, the fabrication process of small planar detectors at USD is discussed in great detail. The impact of the procedure and geometry on the detector performance was analyzed for eight detectors. We characterized the detectors by measuring the leakage current, capacitance, and energy resolution at 662 keV with a Cs-137 source. Four detectors show good performance, which indicates that crystals grown at USD are suitable for making HPGe detectors.
79 - R. Panth , J. Liu , I. Abt 2020
For the first time, planar high-purity germanium detectors with thin amorphous germanium contacts were successfully operated directly in liquid nitrogen and liquid argon in a cryostat at the Max-Planck-Institut fuer Physics in Munich. The detectors were fabricated at the Lawrence Berkeley National Laboratory and the University of South Dakota, using crystals grown at the University of South Dakota. They survived long-distance transportation and multiple thermal cycles in both cryogenic liquids and showed reasonable leakage currents and spectroscopic performance. Also discussed are the pros and cons of using thin amorphous semiconductor materials as an alternative contact technology in large-scale germanium experiments searching for physics beyond the Standard Model.
74 - O. Rest 2019
The most common method to measure direct current high voltage (HV) down to the ppm-level is to use resistive high-voltage dividers. Such devices scale the HV into a range where it can be compared with precision digital voltmeters to reference voltages sources, which can be traced back to Josephson voltage standards. So far the calibration of the scale factors of HV dividers for voltages above 1~kV could only be done at metrology institutes and sometimes involves round-robin tests among several institutions to get reliable results. Here we present a novel absolute calibration method based on the measurement of a differential scale factor, which can be performed with commercial equipment and outside metrology institutes. We demonstrate that reproducible measurements up to 35~kV can be performed with relative uncertainties below $1cdot10^{-6}$. This method is not restricted to metrology institutes and offers the possibility to determine the linearity of high-voltage dividers for a wide range of applications.
The GERmanium Detector Array, GERDA, will search for neutrinoless double beta decay in 76Ge at the National Gran Sasso Laboratory of the INFN. Bare high-purity germanium detectors enriched in 76Ge will be submerged in liquid argon serving simultaneously as a shield against external radioactivity and as a cooling medium. In GERDA Phase-I, reprocessed enriched-Ge detectors, which were previously operated by the Heidelberg-Moscow and IGEX collaborations, will be redeployed. Before operating the enriched detectors, tests are performed with non-enriched bare HPGe detectors in the GERDA underground Detector Laboratory to test the Phase-I detector assembly, the detector handling protocols, the refurbishment technology and to study the long-term stability in liquid argon. The leakage currents in liquid argon and liquid nitrogen have been extensively studied under varying gamma irradiation conditions. In total three non-enriched high-purity p-type prototype germanium detectors have been operated successfully. The detector performance is stable over the long-term measurements. For the first time, performance of bare high-purity germanium detectors in liquid argon is reported.