We present analytical calculations, Finite Element Analysis modeling, and physical measurements of the interstrip capacitances for different potential strip geometries and dimensions of the readout boards for the GE2/1 triple-Gas Electron Multiplier detector in the CMS muon system upgrade. The main goal of the study is to find configurations that minimize the interstrip capacitances and consequently maximize the signal-to-noise ratio for the detector. We find agreement at the 1.5--4.8% level between the two methods of calculations and on the average at the 17% level between calculations and measurements. A configuration with halved strip lengths and doubled strip widths results in a measured 27--29% reduction over the original configuration while leaving the total number of strips unchanged. We have now adopted this design modification for all eight module types of the GE2/1 detector and will produce the final detector with this new strip design.

قيم البحث

اقرأ أيضاً

The muon detection system of the Compact Muon Solenoid experiment at the CERN Large Hadron Collider is based on different technologies for muon tracking and triggering. In particular, the muon system in the endcap disks of the detector consists of Re sistive Plate Chambers for triggering and Cathode Strip Chambers for tracking. At present, the endcap muon system is only partially instrumented with the very forward detector region remaining uncovered. In view of a possible future extension of the muon endcap system, we report on a feasibility study on the use of Micro-Pattern Gas Detectors, in particular Gas Electron Multipliers, for both muon triggering and tracking. Results on the construction and characterization of small tripleGas Electron Multiplier prototype detectors are presented.
At present, part of the forward RPC muon system of the CMS detector at the CERN LHC remains uninstrumented in the high-eta region. An international collaboration is investigating the possibility of covering the 1.6 < |eta| < 2.4 region of the muon en dcaps with large-area triple-GEM detectors. Given their good spatial resolution, high rate capability, and radiation hardness, these micro-pattern gas detectors are an appealing option for simultaneously enhancing muon tracking and triggering capabilities in a future upgrade of the CMS detector. A general overview of this feasibility study will be presented. The design and construction of small (10times10 cm2) and full-size trapezoidal (1times0.5 m2) triple-GEM prototypes will be described. During detector assembly, different techniques for stretching the GEM foils were tested. Results from measurements with x-rays and from test beam campaigns at the CERN SPS will be shown for the small and large prototypes. Preliminary simulation studies on the expected muon reconstruction and trigger performances of this proposed upgraded muon system will be reported.
119 - M. Posik , B. Surrow 2016
Many experiments are currently using or proposing to use large area GEM foils in their detectors, which is creating a need for commercially available GEM foils. Currently CERN is the only main distributor of large GEM foils, however with the growing interest in GEM technology keeping up with the increasing demand for GEMs will be difficult. Thus the commercialization of GEMs up to 50 $times$ 50 cm$^2$ has been established by Tech-Etch Inc. of Plymouth, MA, USA using the single-mask technique. The electrical performance and optical quality of the single-mask GEM foils have been found to be on par with those produced by CERN. The next critical step towards validating the Tech-Etch single-mask GEM foils is to test their performance under physics conditions. These measurements will allow us to quantify and compare the gain and efficiency of the detector to other triple-GEM detectors. This will be done by constructing several single-mask triple-GEM detectors, using foils manufactured by Tech-Etch, which follow the design used by the STAR Forward GEM Tracker (FGT). These detectors will investigate ways in which to further decrease the material budget and increase the efficiency of the detector by incorporating perforated Kapton spacer rings rather than G10 spacing grids to reduce the dead area of the detector. The materials and tooling needed to assemble the triple-GEM detectors have been acquired. The GEM foils have been electrically tested, and a handful have been optically scanned. We found these results to be consistent with GEM foils produced by CERN. With the success of these initial tests, construction of the triple-GEM detectors is now under way.
Gas Electron Multiplier (GEM) technology is being considered for the forward muon upgrade of the CMS experiment in Phase 2 of the CERN LHC. Its first implementation is planned for the GE1/1 system in the $1.5 < midetamid < 2.2$ region of the muon end cap mainly to control muon level-1 trigger rates after the second long LHC shutdown. A GE1/1 triple-GEM detector is read out by 3,072 radial strips with 455 $mu$rad pitch arranged in eight $eta$-sectors. We assembled a full-size GE1/1 prototype of 1m length at Florida Tech and tested it in 20-120 GeV hadron beams at Fermilab using Ar/CO$_{2}$ 70:30 and the RD51 scalable readout system. Four small GEM detectors with 2-D readout and an average measured azimuthal resolution of 36 $mu$rad provided precise reference tracks. Construction of this largest GEM detector built to-date is described. Strip cluster parameters, detection efficiency, and spatial resolution are studied with position and high voltage scans. The plateau detection efficiency is [97.1 $pm$ 0.2 (stat)]%. The azimuthal resolution is found to be [123.5 $pm$ 1.6 (stat)] $mu$rad when operating in the center of the efficiency plateau and using full pulse height information. The resolution can be slightly improved by $sim$ 10 $mu$rad when correcting for the bias due to discrete readout strips. The CMS upgrade design calls for readout electronics with binary hit output. When strip clusters are formed correspondingly without charge-weighting and with fixed hit thresholds, a position resolution of [136.8 $pm$ 2.5 stat] $mu$rad is measured, consistent with the expected resolution of strip-pitch/$sqrt{12}$ = 131.3 $mu$rad. Other $eta$-sectors of the detector show similar response and performance.
The LHC is undergoing a high luminosity upgrade, which is set to increase the instantaneous luminosity by at least a factor of five, resulting in a higher muon flux rate in the forward region, which will overwhelm the current trigger system of the CM S experiment. The ME0, a gas electron multiplier detector, is proposed for the Phase-2 Muon System Upgrade to help increase the muon acceptance and to control the Level 1 muon trigger rate. To lower the probability of HV discharges, the ME0 was designed with GEM foils that are segmented on both sides. Initial testing of the ME0 showed substantial crosstalk between readout sectors. Here, we investigate, characterize, and quantify the crosstalk in the detector, and estimate the performance of the chamber as a result of this crosstalk via simulation of the detector dead time, efficiency loss, and frontend electronics response. The results of crosstalk via signals produced by applying a square voltage pulse directly on the readout strips of the detector with a pulser are summarized, and the efficacy of various mitigation strategies are presented. The crosstalk is a result of capacitive coupling between the readout strips on the readout board and between the readout strips and the bottom of GEM3. The crosstalk also generally follows a pattern where the largest magnitude of crosstalk is within the same azimuthal readout segment in the detector and in the nearest horizontal segments. The use of bypass capacitors and larger HV segments successfully reduce the crosstalk: we observe a maximum decrease of crosstalk in sectors previously experiencing crosstalk from $(1.66pm0.03)%$ to $(1.11pm0.02)%$ with all HV segments connected in parallel on the bottom of GEM3, with an HV low-pass filter, and an HV divider. These mitigation strategies slightly increase crosstalk $big(hspace{-0.1cm}lessapprox 0.4%big)$ in readout sectors farther away.