ترغب بنشر مسار تعليمي؟ اضغط هنا

A New Proton CT Scanner

68   0   0.0 ( 0 )
 نشر من قبل Sergey Uzunyan
 تاريخ النشر 2014
  مجال البحث فيزياء
والبحث باللغة English
 تأليف S. A. Uzunyan




اسأل ChatGPT حول البحث

The design, construction, and preliminary testing of a second generation proton CT scanner is presented. All current treatment planning systems at proton therapy centers use X-ray CT as the primary imaging modality for treatment planning to calculate doses to tumor and healthy tissues. One of the limitations of X-ray CT is in the conversion of X-ray attenuation coefficients to relative (proton) stopping powers, or RSP. This results in more proton range uncertainty, larger target volumes and therefore, more dose to healthy tissues. To help improve this, we present a novel scanner capable of high dose rates, up to 2~MHz, and large area coverage, 20~x~24~cm$^2$, for imaging an adult head phantom and reconstructing more accurate RSP values.

قيم البحث

اقرأ أيضاً

79 - S. A. Uzunyan 2013
We describe the development of a proton Computed Tomography (pCT) scanner at Northern Illinois University (NIU) in collaboration with Fermilab and Delhi University. This paper provides an overview of major components of the scanner and a detailed description of the data acquisition system (DAQ).
The TT-PET collaboration is developing an MRI-compatible small animal PET scanner in which the sensitive element is a monolithic silicon pixel ASIC targeting 30 ps RMS time resolution. The photon-detection technique is based on a stack of alternating layers of high-Z photon converter and 100 $mathrm{mu m}$ silicon sensors, to produce a scanner with 0.5 $mathrm{times}$ 0.5 $mathrm{times}$ 0.2 $mathrm{mm^{3}}$ granularity for precise depth-of-interaction measurement. In this paper we present the results of simulation studies for the expected data rate, time-of-flight and spatial resolution, as well as the performance of image reconstruction with and without the use of timing information.
The purpose of the presented research is estimation of the performance characteristics of the economic Total-Body Jagiellonian-PET system (TB-J-PET) constructed from plastic scintillators. The characteristics are estimated according to the NEMA NU-2- 2018 standards utilizing the GATE package. The simulated detector consists of 24 modules, each built out of 32 plastic scintillator strips (each with cross section of 6 mm times 30 mm and length of 140 cm or 200 cm) arranged in two layers in regular 24-sided polygon circumscribing a circle with the diameter of 78.6 cm. For the TB-J-PET with an axial field-of-view (AFOV) of 200 cm, a spatial resolutions of 3.7 mm (transversal) and 4.9 mm (axial) are achieved. The NECR peak of 630 kcps is expected at 30 kBq/cc activity concentration and the sensitivity at the center amounts to 38 cps/kBq. The SF is estimated to 36.2 %. The values of SF and spatial resolution are comparable to those obtained for the state-of-the-art clinical PET scanners and the first total-body tomographs: uExplorer and PennPET. With respect to the standard PET systems with AFOV in the range from 16 cm to 26 cm, the TB-J-PET is characterized by an increase in NECR approximately by factor of 4 and by the increase of the whole-body sensitivity by factor of 12.6 to 38. The TOF resolution for the TB-J-PET is expected to be at the level of CRT=240 ps (FWHM). For the TB-J-PET with an axial field-of-view (AFOV) of 140 cm, an image quality of the reconstructed images of a NEMA IEC phantom was presented with a contrast recovery coefficient (CRC) and a background variability parameters. The increase of the whole-body sensitivity and NECR estimated for the TB-J-PET with respect to current commercial PET systems makes the TB-J-PET a promising cost-effective solution for the broad clinical applications of total-body PET scanners.
Time-Over-Threshold (TOT) technique is being used widely due to its implications in developing the multi channel readouts mainly when fast signal processing is required. Using TOT technique as a measure of energy loss instead of charge integration me thods significantly reduces the signals readout cost by combining the time and energy information. Therefore, this approach can potentially be used in J-PET tomograph which is build from plastic scintillators characterized by fast light signals. The drawback in adopting this technique is lying in the non-linear correlation between input energy loss and TOT of the signal. The main motivation behind this work is to develop the relationship between TOT and energy loss and validate it with the J-PET tomograph. The experiment was performed using the $^{22}$Na beta emitter source placed in the center of the J-PET tomograph. One can obtain primary photons of two different energies: 511 keV photon from the annihilation of positron (direct annihilation or through the formation of para-Positronim atom or pick-off process of ortho-Positronium atoms), and 1275 keV prompt photon. This allows to study the correlation between TOT values and energy loss for energy range up to 1000 keV. As the photon interacts dominantly via Compton scattering inside the plastic scintillator, there is no direct information of primary photon energy. However, using the J-PET geometry one can measure the scattering angle of the interacting photon. Since, $^{22}$Na source emits photons of two different energies, it is required to know unambiguously the energy of incident photons and its corresponding scattering angle for the estimation of energy deposition. In this work, the relationship between Time Over Threshold and energy loss by interacting photons inside the plastic scintillators used in J-PET scanner is established for a energy deposited range 100-1000 keV
We propose an innovative method for proton radiography based on nuclear emulsion film detectors, a technique in which images are obtained by measuring the position and the residual range of protons passing through the patients body. For this purpose, nuclear emulsion films interleaved with tissue equivalent absorbers can be used to reconstruct proton tracks with very high accuracy. This is performed through a fully automated scanning procedure employing optical microscopy, routinely used in neutrino physics experiments. Proton radiography can be used in proton therapy to obtain direct information on the average tissue density for treatment planning optimization and to perform imaging with very low dose to the patient. The first prototype of a nuclear emulsion based detector has been conceived, constructed and tested with a therapeutic proton beam. The first promising experimental results have been obtained by imaging simple phantoms.
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا