ﻻ يوجد ملخص باللغة العربية
The International Electrotechnical Commission (IEC) has previously defined standard rotation operators for positive gantry, collimator and couch rotations for the radiotherapy DICOM coordinate system that is commonly used by treatment planning systems. Coordinate transformations to the coordinate systems of commonly used Monte Carlo (MC) codes (BEAMnrc/DOSXYZnrc and VMC++) have been derived and published in the literature. However, these coordinate transformations disregard patient orientation during the computed tomography (CT) scan, and assume the most commonly used head first, supine orientation. While less common, other patient orientations are used in clinics - Monte Carlo verification of such treatments can be problematic due to the lack of appropriate coordinate transformations. In this work, a solution has been obtained by correcting the CT-derived phantom orientation and deriving generalized coordinate transformations for field angles in the DOSXYZnrc and VMC++ codes. The rotation operator that includes any possible patient treatment orientation was determined using the DICOM Image Orientation tag (0020,0037). The derived transformations of the patient image and beam direction angles were verified by comparison of MC dose distributions with the Eclipse treatment planning system, calculated for each of the eight possible patient orientations.
Monte Carlo (MC) methods provide the most accurate to-date dose calculations in heterogeneous media and complex geometries, and this spawns increasing interest in incorporating MC calculations into treatment planning quality assurance process. This i
We have previously described RapidBrachyMCTPS, a brachytherapy treatment planning toolkit consisting of a graphical user interface (GUI) and a Geant4-based Monte Carlo (MC) dose calculation engine. This work describes the tools that have recently bee
Purpose: A Monte Carlo (MC) beam model and its implementation in a clinical treatment planning system (TPS, Varian Eclipse) are presented for a modified ultra-high dose-rate electron FLASH radiotherapy (eFLASH-RT) LINAC. Methods: The gantry head wi
The next great leap toward improving treatment of cancer with radiation will require the combined use of online adaptive and magnetic resonance guided radiation therapy techniques with automatic X-ray beam orientation selection. Unfortunately, by uni
Purpose: To investigate the validity of two Monte Carlo simulation absolute dosimetry approaches in the case of a small field dedicated `D-shaped collimator used for the retinoblastoma treatment with external photon beam radiotherapy. Methods: The