We employed two independent methods to study possible damage to the scintillation mechanism in lead tungstate crystals due to irradiation by a 34 GeV pion beam. First, 10 crystals were irradiated simultaneously over 30 hours by a narrow beam, so that only a small region of each crystal was affected. We studied the effect of the irradiation on the light output non-uniformity. If a localized degradation was observed, it would indicate damage to the scintillation mechanism. Secondly, we detected light output using two phototubes attached to sides of a crystal. Since these phototubes detect scintillation light only from a small localized region, the effect of transmission loss should be minimal. We did not see any statistically significant evidence for scintillation mechanism damage with either method. The effect is consistent with zero, and the upper limit is 0.5% at 95% C.L.
Studies of the radiation hardness of lead tungstate crystals produced by the Bogoroditsk Techno-Chemical Plant in Russia and the Shanghai Institute of Ceramics in China have been carried out at IHEP, Protvino. The crystals were irradiated by a 40-GeV
Ensuring the radiation hardness of PbWO4 crystals was one of the main priorities during the construction of the electromagnetic calorimeter of the CMS experiment at CERN. The production on an industrial scale of radiation hard crystals and their cert
A Cerium Fluoride crystal produced during early R&D studies for calorimetry at the CERN Large Hadron Collider was exposed to a 24 GeV/c proton fluence Phi_p=(2.78 +- 0.20) x 10EE13 cm-2 and, after one year of measurements tracking its recovery, to a
Fast hadrons have been observed to cause a cumulative damage in Lead Tungstate and LYSO crystals. The underlying mechanism has been proven to be the creation of fission tracks, which act as scattering centres, thus reducing the light collection effic
This paper is devoted to the study of a degradation of CsI(Tl)crystals scintillation characteristics under irradiation with gamma-quanta at the uniformly distributed absorbed dose up to 3700 rad. The sample set consisted of 25 crystals of 30 cm long