ترغب بنشر مسار تعليمي؟ اضغط هنا

A new geometry of scintillating crystals with Strip SiPMs: a PET detector with precise position and time determination

142   0   0.0 ( 0 )
 نشر من قبل Zheng Liu
 تاريخ النشر 2019
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

Measurement of the Time-of-Flight (TOF) of the 511 keV gammas brings an important reduction of statistical noise in the PET image, with higher precision time measurements producing clearer images. Scintillating crystals are used to convert the 511 keV annihilation photon to an electron of ~511 KeV energy via the photoelectric effect; it is necessary to determine with precision the position and time of this conversion within the scintillating crystal. We propose using an array of crystals cut into a specific geometry discussed below; these crystals are read out by an array of strip SiPMs. This technique allows individual time measurements of the first arriving photo-electrons and to extract the best time resolution using a specific algorithm. The final result is a precise determination of the 3D position (that includes the depth of interaction) of the photoelectric interaction and an improved time measurement.

قيم البحث

اقرأ أيضاً

60 - A. Fedorov 2002
Crystal arrays made of LSO and LuAP crystals 2x2x10 mm pixels were manufactured for evaluation of detector with depth-of-interaction (DOI) determination capability intended for small animal positron emission tomograph. Position-sensitive LSO/LuAP pho swich DOI detector based on crystal 8x8 arrays and HAMAMATSU R5900-00-M64 position-sensitive multi-anode photomultiplier tube was developed and evaluated. Time resolution was found to be not worse than 1.0 ns FWHM for both layers, and spatial resolution mean value was 1.5 mm FWHM for the center of field-of-view.
The fast scintillation decay time and the high scintillation yield of liquid xenon makes it an appropriate material for nuclear medicine. Moreover, being a continuous medium with a uniform response, liquid xenon allows one to avoid most of the geomet rical distortions of conventional detectors based on scintillating crystals. In this paper, we describe how these properties have motivated the development of a novel concept for positron emission tomography scanners with Time-Of-Flight measurement, which uses liquid xenon as a scintillating material and silicon photomultipliers as sensors. Monte Carlo studies have indicated that this technology would provide a very good intrinsic time resolution, of around 70 ps. Moreover, being liquid xenon transparent to UV and blue wavelengths, both scintillation and Cherenkov light can be exploited. While the former can be used for energy measurements, the latter is a prompt signal (of a few picoseconds), which provides a very precise time measurement. Monte Carlo simulations point to a time resolution of 30-50 ps obtained using Cherenkov light. A first prototype is being built to demonstrate the high energy, spatial and time resolution of this concept, using a ring of around 30 cm of internal diameter and a depth of 3 cm instrumented with VUV--sensitive silicon photomultipliers.
A photon-counting silicon strip detector with two energy thresholds was investigated for spectral X-ray imaging in a mammography system. Preliminary studies already indicate clinical benefit of the detector, and the purpose of the present study is op timization with respect to energy resolution. Factors relevant for the energy response were measured, simulated, or gathered from previous studies, and used as input parameters to a cascaded detector model. Threshold scans over several X-ray spectra were used to calibrate threshold levels to energy, and to validate the model. The energy resolution of the detector assembly was assessed to range over DeltaE/E = 0.12-0.26 in the mammography region. Electronic noise dominated the peak broadening, followed by charge sharing between adjacent detector strips, and a channel-to-channel threshold spread. The energy resolution may be improved substantially if these effects are reduced to a minimum. Anti-coincidence logic mitigated double counting from charge sharing, but erased the energy resolution of all detected events, and optimization of the logic is desirable. Pile-up was found to be of minor importance at typical mammography rates.
This paper describes the characterization of crystal matrices and silicon photomultiplier arrays for a novel Positron Emission Tomography (PET) detector, namely the external plate of the EndoTOFPET-US system. The EndoTOFPET-US collaboration aims to i ntegrate Time-Of-Flight PET with ultrasound endoscopy in a novel multimodal device, capable to support the development of new biomarkers for prostate and pancreatic tumors. The detector consists in two parts: a PET head mounted on an ultrasound probe and an external PET plate. The challenging goal of 1 mm spatial resolution for the PET image requires a detector with small crystal size, and therefore high channel density: 4096 LYSO crystals individually readout by Silicon Photomultipliers (SiPM) make up the external plate. The quality and properties of these components must be assessed before the assembly. The dark count rate, gain, breakdown voltage and correlated noise of the SiPMs are measured, while the LYSO crystals are evaluated in terms of light yield and energy resolution. In order to effectively reduce the noise in the PET image, high time resolution for the gamma detection is mandatory. The Coincidence Time Resolution (CTR) of all the SiPMs assembled with crystals is measured, and results show a value close to the demanding goal of 200 ps FWHM. The light output is evaluated for every channel for a preliminary detector calibration, showing an average of about 1800 pixels fired on the SiPM for a 511 keV interaction. Finally, the average energy resolution at 511 keV is about 13 %, enough for effective Compton rejection.
114 - P. Moskal , O. Rundel , D. Alfs 2016
Recent tests of a single module of the Jagiellonian Positron Emission Tomography system (J-PET) consisting of 30 cm long plastic scintillator strips have proven its applicability for the detection of annihilation quanta (0.511 MeV) with a coincidence resolving time (CRT) of 0.266 ns. The achieved resolution is almost by a factor of two better with respect to the current TOF-PET detectors and it can still be improved since, as it is shown in this article, the intrinsic limit of time resolution for the determination of time of the interaction of 0.511 MeV gamma quanta in plastic scintillators is much lower. As the major point of the article, a method allowing to record timestamps of several photons, at two ends of the scintillator strip, by means of matrix of silicon photomultipliers (SiPM) is introduced. As a result of simulations, conducted with the number of SiPM varying from 4 to 42, it is shown that the improvement of timing resolution saturates with the growing number of photomultipliers, and that the 2 x 5 configuration at two ends allowing to read twenty timestamps, constitutes an optimal solution. The conducted simulations accounted for the emission time distribution, photon transport and absorption inside the scintillator, as well as quantum efficiency and transit time spread of photosensors, and were checked based on the experimental results. Application of the 2 x 5 matrix of SiPM allows for achieving the coincidence resolving time in positron emission tomography of $approx$ 0.170 ns for 15 cm axial field-of-view (AFOV) and $approx$ 0.365 ns for 100 cm AFOV. The results open perspectives for construction of a cost-effective TOF-PET scanner with significantly better TOF resolution and larger AFOV with respect to the current TOF-PET modalities.
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا