ترغب بنشر مسار تعليمي؟ اضغط هنا

In-flight calibration and verification of the Planck-LFI instrument

160   0   0.0 ( 0 )
 نشر من قبل Anna Gregorio Dr.
 تاريخ النشر 2013
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

In this paper we discuss the Planck-LFI in-flight calibration campaign. After a brief overview of the ground test campaigns, we describe in detail the calibration and performance verification (CPV) phase, carried out in space during and just after the cool-down of LFI. We discuss in detail the functionality verification, the tuning of the front-end and warm electronics, the preliminary performance assessment and the thermal susceptibility tests. The logic, sequence, goals and results of the in-flight tests are discussed. All the calibration activities were successfully carried out and the instrument response was comparable to the one observed on ground. For some channels the in-flight tuning activity allowed us to improve significantly the noise performance.



قيم البحث

اقرأ أيضاً

The Low Frequency Instrument on board the PLANCK satellite is designed to give the most accurate map ever of the CMB anisotropy of the whole sky over a broad frequency band spanning 27 to 77 GHz. It is made of an array of 22 pseudo-correlation radiom eters, composed of 11 actively cooled (20 K) Front End Modules (FEMs), and 11 Back End Modules (BEMs) at 300K. The connection between the two parts is made with rectangular Wave Guides. Considerations of different nature (thermal, electromagnetic and mechanical), imposed stringent requirements on the WGs characteristics and drove their design. From the thermal point of view, the WG should guarantee good insulation between the FEM and the BEM sections to avoid overloading the cryocooler. On the other hand it is essential that the signals do not undergo excessive attenuation through the WG. Finally, given the different positions of the FEM modules behind the focal surface and the mechanical constraints given by the surrounding structures, different mechanical designs were necessary. A composite configuration of Stainless Steel and Copper was selected to satisfy all the requirements. Given the complex shape and the considerable length (about 1.5-2 m), manufacturing and testing the WGs was a challenge. This work deals with the development of the LFI WGs, including the choice of the final configuration and of the fabrication process. It also describes the testing procedure adopted to fully characterize these components from the electromagnetic point of view and the space qualification process they underwent. Results obtained during the test campaign are reported and compared with the stringent requirements. The performance of the LFI WGs is in line with requirements, and the WGs were successfully space qualified.
this paper is part of the Prelaunch status LFI papers published on JINST: http://www.iop.org/EJ/journal/-page=extra.proc5/jinst The Low Frequency Instrument is optically interfaced with the ESA Planck telescope through 11 corrugated feed horns each c onnected to the Radiometer Chain Assembly (RCA). This paper describes the design, the manufacturing and the testing of the flight model feed horns. They have been designed to optimize the LFI optical interfaces taking into account the tight mechanical requirements imposed by the Planck focal plane layout. All the eleven units have been successfully tested and integrated with the Ortho Mode transducers.
The Low Frequency Instrument (LFI) of the ESA Planck CMB mission is an array of 22 ultra sensitive pseudocorrelation radiometers working at 30, 44, and 70 GHz. LFI has been calibrated and delivered for integration with the satellite to the European S pace Agency on November 2006. The aim of Planck is to measure the anisotropy and polarization of the Cosmic Background Radiation with a sensitivity and angular resolution never reached before over the full sky. LFI is intrinsically sensitive to polarization thanks to the use of Ortho-Mode Transducers (OMT) located between the feedhorns and the pseudo-correlation radiometers. The OMTs are microwave passive components that divide the incoming radiation into two linear orthogonal components. A set of 11 OMTs (2 at 30 GHz, 3 at 44 GHz, and 6 at 70 GHz) were produced and tested. This work describes the design, development and performance of the eleven Flight Model OMTs of LFI. The final design was reached after several years of development. At first, Elegant Bread Board OMTs were produced to investigate the manufacturing technology and design requirements. Then, a set of 3 Qualification Model (QM) OMTs were designed, manufactured and tested in order to freeze the design and the manufacturing technology for the flight units. Finally, the Flight Models were produced and tested. It is shown that all the OMT units have been accepted for flight and the electromagnetic performance is at least marginally compliant with the requirements. Mechanically, the units passed all the thermoelastic qualification tests after a reworking necessary after the QM campaign.
250 - F. Villa , L. Terenzi , M. Sandri 2010
The Low Frequency Instrument (LFI) on-board the ESA Planck satellite carries eleven radiometer subsystems, called Radiometer Chain Assemblies (RCAs), each composed of a pair of pseudo-correlation receivers. We describe the on-ground calibration campa ign performed to qualify the flight model RCAs and to measure their pre-launch performances. Each RCA was calibrated in a dedicated flight-like cryogenic environment with the radiometer front-end cooled to 20K and the back-end at 300K, and with an external input load cooled to 4K. A matched load simulating a blackbody at different temperatures was placed in front of the sky horn to derive basic radiometer properties such as noise temperature, gain, and noise performance, e.g. 1/f noise. The spectral response of each detector was measured as was their susceptibility to thermal variation. All eleven LFI RCAs were calibrated. Instrumental parameters measured in these tests, such as noise temperature, bandwidth, radiometer isolation, and linearity, provide essential inputs to the Planck-LFI data analysis.
We discuss the methods employed to photometrically calibrate the data acquired by the Low Frequency Instrument on Planck. Our calibration is based on a combination of the Orbital Dipole plus the Solar Dipole, caused respectively by the motion of the Planck spacecraft with respect to the Sun and by motion of the Solar System with respect to the CMB rest frame. The latter provides a signal of a few mK with the same spectrum as the CMB anisotropies and is visible throughout the mission. In this data release we rely on the characterization of the Solar Dipole as measured by WMAP. We also present preliminary results (at 44GHz only) on the study of the Orbital Dipole, which agree with the WMAP value of the Solar System speed within our uncertainties. We compute the calibration constant for each radiometer roughly once per hour, in order to keep track of changes in the detectors gain. Since non-idealities in the optical response of the beams proved to be important, we implemented a fast convolution algorithm which considers the full beam response in estimating the signal generated by the dipole. Moreover, in order to further reduce the impact of residual systematics due to sidelobes, we estimated time variations in the calibration constant of the 30GHz radiometers (the ones with the largest sidelobes) using the signal of a reference load. We have estimated the calibration accuracy in two ways: we have run a set of simulations to assess the impact of statistical errors and systematic effects in the instrument and in the calibration procedure, and we have performed a number of consistency checks on the data and on the brightness temperature of Jupiter. Calibration errors for this data release are expected to be about 0.6% at 44 and 70 GHz, and 0.8% at 30 GHz. (Abriged.)
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا