Flux dependent non-linearity (reciprocity failure) in HgCdTe near infrared detectors can severely impact an instruments performance, in particular with respect to precision photometric measurements. The cause of this effect is presently not understood. To investigate reciprocity failure, a dedicated test system was built. For flux levels between 1 and 50,000 photons/s, a sensitivity to reciprocity failure of approximately 0.1%/decade was achieved. A wavelength independent non-linearity due to reciprocity failure of about 0.35%/decade was measured in a 1.7 micron HgCdTe detector.
A detailed study of reciprocity failure in four 1.7 micron cutoff HgCdTe near-infrared detectors is presented. The sensitivity to reciprocity failure is approximately 0.1%decade over up to five orders of magnitude in illumination intensity. The four
Flux dependent non-linearity (reciprocity failure) in HgCdTe NIR detectors with 1.7 micron cut-off was investigated. A dedicated test station was designed and built to measure reciprocity failure over the full dynamic range of near infrared detectors
Weak gravitational lensing has emerged as a leading probe of the growth of cosmic structure. However, the shear signal is very small and accurate measurement depends critically on our ability to understand how non-ideal instrumental effects affect as
We study the nonlinearity (NL) in the conversion from charge to voltage in infrared detectors (HXRG) for use in precision astronomy. We present laboratory measurements of the NL function of a H2RG detector and discuss the accuracy to which it would n
High-precision astrometry requires accurate point-spread function modeling and accurate geometric-distortion corrections. This paper demonstrates that it is possible to achieve both requirements with data collected at the high acuity wide-field K-ban