In this paper, we report on the design and operation of the LongBo time projection chamber in the Liquid Argon Purity Demonstrator cryostat. This chamber features a 2 m long drift distance. We measure the electron drift lifetime in the liquid argon u sing cosmic ray muons and the lifetime is at least 14 ms at 95% confidence level. LongBo is equipped with preamplifiers mounted on the detector in the liquid argon. Of the 144 channels, 128 channels were readout by preamplifiers made with discrete circuitry and 16 channels were readout by ASIC preamplifiers. For the discrete channels, we measure a signal-to-noise (S/N) ratio of 30 at a drift field of 350 V/cm. The measured S/N ratio for the ASIC channels was 1.4 times larger than that measured for the discrete channels.
We trap cold, ground-state, argon atoms in a deep optical dipole trap produced by a build-up cavity. The atoms, which are a general source for the sympathetic cooling of molecules, are loaded in the trap by quenching them from a cloud of laser-cooled metastable argon atoms. Although the ground state atoms cannot be directly probed, we detect them by observing the collisional loss of co-trapped metastable argon atoms using a new type of parametric loss spectroscopy. Using this technique we also determine the polarizability ratio between the ground and the metastable 4s[3/2]$_2$ state to be 40$pm6$ and find a polarisability of (7.3$pm$1.1) $times$10$^{-39}$ Cm$^2/$V for the metastable state. Finally, Penning and associative losses of metastable atoms, in the absence of light assisted collisions, are determined to be $(3.3pm 0.8) times 10^{-10}$ cm$^3$s$^{-1}$.