The electronic spin of the nitrogen vacancy (NV) center in diamond forms an atomically sized, highly sensitive sensor for magnetic fields. To harness the full potential of individual NV centers for sensing with high sensitivity and nanoscale spatial resolution, NV centers have to be incorporated into scanning probe structures enabling controlled scanning in close proximity to the sample surface. Here, we present an optimized procedure to fabricate single-crystal, all-diamond scanning probes starting from commercially available diamond and show a highly efficient and robust approach for integrating these devices in a generic atomic force microscope. Our scanning probes consisting of a scanning nanopillar (200 nm diameter, $1-2,mu$m length) on a thin ($< 1mu$m) cantilever structure, enable efficient light extraction from diamond in combination with a high magnetic field sensitivity ($mathrm{eta_{AC}}approx50pm20,mathrm{nT}/sqrt{mathrm{Hz}}$). As a first application of our scanning probes, we image the magnetic stray field of a single Ni nanorod. We show that this stray field can be approximated by a single dipole and estimate the NV-to-sample distance to a few tens of nanometer, which sets the achievable resolution of our scanning probes.
The negatively-charged nitrogen-vacancy center (NV) in diamond forms a versatile system for quantum sensing applications. Combining the advantageous properties of this atomic-sized defect with scanning probe techniques such as atomic force microscopy (AFM) enables nanoscale imaging of e.g. magnetic fields. To form a scanning probe device, we place single NVs shallowly (i.e. < 20 nm) below the top facet of a diamond nanopillar, which is located on a thin diamond platform of typically below 1 mu m thickness. This device can be attached to an AFM head, forming an excellent scanning probe tip. Furthermore, it simultaneously influences the collectible photoluminescence (PL) rate of the NV located inside. Especially sensing protocols using continuous optically-detected magnetic resonance (ODMR) benefit from an enhanced collectible PL rate, improving the achievable sensitivity. This work presents a comprehensive set of simulations to quantify the influence of the device geometry on the collectible PL rate for individual NVs. Besides geometric parameters (e.g. pillar length, diameter and platform thickness), we also focus on fabrication uncertainties such as the exact position of the NV or the taper geometry of the pillar introduced by imperfect etching. As a last step, we use these individual results to optimize our current device geometry, yielding a realistic gain in collectible PL rate by a factor of 13 compared to bulk diamond and 1.8 compared to our unoptimized devices.
Scanning Thermal Microscopy (SThM) uses micromachined thermal sensors integrated in a force sensing cantilever with a nanoscale tip can be highly useful for exploration of thermal management of nanoscale semiconductor devices. As well as mapping of surface properties of related materials. Whereas SThM is capable to image externally generated heat with nanoscale resolution, its ability to map and measure thermal conductivity of materials has been mainly limited to polymers or similar materials possessing low thermal conductivity in the range from 0.1 to 1 W/mK, with lateral resolution on the order of 1 mum. In this paper we use linked experimental and theoretical approaches to analyse thermal performance and sensitivity of the micromachined SThM probes in order to expand their applicability to a broader range of nanostructures from polymers to semiconductors and metals. We develop physical models of interlinked thermal and electrical phenomena in these probes and then validate these models using experimental measurements of the real probes, which provided the basis for analysing SThM performance in exploration of nanostructures. Our study then highlights critical features of these probes, namely, the geometrical location of the thermal sensor with respect to the probe apex, thermal conductance of the probe to the support base, heat conduction to the surrounding gas, and the thermal conductivity of tip material adjacent to the apex. It is furthermore allows us to propose a novel design of the SThM probe that incorporates a carbon nanotube (CNT) or similar high thermal conductivity graphene sheet material positioned near the probe apex. The new sensor is predicted to provide spatial resolution to the thermal properties of nanostructures on the order of few tens of nm, as well as to expand the sensitivity of the SThM probe to materials with heat conductivity values up to 100-1000 W/mK.
Scanning diamond magnetometers based on the optically detected magnetic resonance of the nitrogen-vacancy centre offer very high sensitivity and non-invasive imaging capabilities when the stray fields emanating from ultrathin magnetic materials are sufficiently low (< 10 mT). Beyond this low-field regime, the optical signal quenches and a quantitative measurement is challenging. While the field-dependent NV photoluminescence can still provide qualitative information on magnetic morphology, this operation regime remains unexplored particularly for surface magnetisation larger than $sim$ 3 mA. Here, we introduce a multi-angle reconstruction technique (MARe) that captures the full nanoscale domain morphology in all magnetic-field regimes leading to NV photoluminescence quench. To demonstrate this, we use [Ir/Co/Pt]$_{14}$ multilayer films with surface magnetisation an order of magnitude larger than previous reports. Our approach brings non-invasive nanoscale magnetic field imaging capability to the study of a wider pool of magnetic materials and phenomena.
Detection of AC magnetic fields at the nanoscale is critical in applications ranging from fundamental physics to materials science. Isolated quantum spin defects, such as the nitrogen-vacancy center in diamond, can achieve the desired spatial resolution with high sensitivity. Still, vector AC magnetometry currently relies on using different orientations of an ensemble of sensors, with degraded spatial resolution, and a protocol based on a single NV is lacking. Here we propose and experimentally demonstrate a protocol that exploits a single NV to reconstruct the vectorial components of an AC magnetic field by tuning a continuous driving to distinct resonance conditions. We map the spatial distribution of an AC field generated by a copper wire on the surface of the diamond. The proposed protocol combines high sensitivity, broad dynamic range, and sensitivity to both coherent and stochastic signals, with broad applications in condensed matter physics, such as probing spin fluctuations.
We present an implementation of all-diamond scanning probes for scanning nitrogen-vacancy (NV) magnetometry fabricated from (111)-oriented diamond material. The realized scanning probe tips on average contain single NV spins, a quarter of which have their spin quantization axis aligned parallel to the tip direction. Such tips enable single-axis vector magnetic field imaging with nanoscale resolution, where the measurement axis is oriented normal to the scan plane. We discuss how these tips bring multiple practical advantages for NV magnetometry, in particular regarding quantitative analysis of the resulting data. We further demonstrate the beneficial optical properties of NVs oriented along the tip direction, such as polarization-insensitive excitation, which simplifies optical setups needed for NV magnetometry. Our results will be impactful for scanning NV magnetometry in general and for applications in spintronics and the investigation of thin film magnets in particular.