The Multi-Blade is a Boron-10-based gaseous detector developed for neutron reflectometry instruments at the European Spallation Source (ESS) in Sweden. The main challenges for neutron reflectometry detectors are the instantaneous counting rate and spatial resolution. The Multi-Blade has been tested on the CRISP reflectometer at the ISIS neutron and muon source in UK. A campaign of scientific measurements has been performed to study the Multi-Blade response in real instrumental conditions. The results of these tests are discussed in this manuscript.
The Multi-Blade is a Boron-10-based gaseous thermal neutron detector developed to face the challenge arising in neutron reflectometry at neutron sources. Neutron reflectometers are challenging instruments in terms of instantaneous counting rate and spatial resolution. This detector has been designed according to the requirements given by the reflectometers at the European Spallation Source (ESS) in Sweden. The Multi-Blade has been installed and tested on the CRISP reflectometer at the ISIS neutron and muon source in UK. The results on the detailed detector characterization are discussed in this manuscript.
The Multi-Blade is a Boron-10-based neutron detector designed for neutron reflectometers and developed for the two instruments (Estia and FREIA) planned for the European Spallation Source in Sweden. A reflectometry demonstrator has been installed at the AMOR reflectometer at the Paul Scherrer Institut (PSI - Switzerland). The setup exploits the Selene guide concept and it can be considered a scaled-down demonstrator of Estia. The results of these tests are discussed. It will be shown how the characteristics of the Multi-Blade detector are features that allow the focusing reflectometry operation mode. Additionally the performance of the Multi-Blade, in terms of rate capability, exceeds current state-of-the-art technology. The improvements with respect to the previous prototypes are also highlighted; from background considerations to the linear and angular uniformity response of the detector.
The European Spallation Source (ESS) is the worlds next generation spallation-based neutron source. The research conducted at ESS will yield in the discovery and development of new materials including the fields of manufacturing, pharmaceuticals, aerospace, engines, plastics, energy, telecommunications, transportation, information technology and biotechnology. The spallation source will deliver an unprecedented neutron flux. In particular, the reflectometers selected for construction, ESTIA and FREIA, have to fulfill challenging requirements. Local incident peak rate can reach 10$^5$~Hz/mm$^2$. For new science to be addressed, the spatial resolution is aimed to be less than 1 mm with a desired scattering of 10$^{-4}$ (peak-to-tail ratio). The latter requirement is approximately two orders of magnitude better than the current state-of-the-art detectors. The main aim of this work is to quantify the cumulative contribution of various detector components to the scattering of neutrons and to prove that the respective effect is within the requirements set for the Multi-Blade detector by the ESS reflectometers. To this end, different sets of geometry and beam parameters are investigated, with primary focus on the cathode coating and the detector window thickness.
We present the results of the measurements of the detection efficiency for a 4.7 r{A} neutron beam incident upon a detector incorporating a stack of up to five MultiWire Proportional Counters (MWPC) with Boron-coated cathodes. The cathodes were made of Aluminum and had a surface exhibiting millimeter-deep V-shaped grooves of 45{deg}, upon which the thin Boron film was deposited by DC magnetron sputtering. The incident neutrons interacting with the converter layer deposited on the sidewalls of the grooves have a higher capture probability, owing to the larger effective absorption film thickness. This leads to a higher overall detection efficiency for the grooved cathode when compared to a cathode with a flat surface. Both the experimental results and the predictions of the GEANT4 model suggests that a 5-counter detector stack with coated grooved cathodes has the same efficiency as a 7-counter stack with flat cathodes. The reduction in the number of counters in the stack without altering the detection efficiency will prove highly beneficial for large-area position-sensitive detectors for neutron scattering applications, for which the cost-effective manufacturing of the detector and associated readout electronics is an important objective. The proposed detector concept could be a technological option for one of the new chopper spectrometers and other instruments planned to be built at the future European Spallation Source in Sweden. These results with macrostructured cathodes generally apply not just to MWPCs but to other gaseous detectors as well.
A Monte Carlo simulation-based optimization of a multilayer 10B-RPC thermal neutron detector is performed targeting an increase in the counting rate capability while maintaining high (>50%) detection efficiency for thermal neutrons. The converter layer thicknesses of individual RPCs are optimized for several configurations of a detector containing a stack of 10 double gap RPCs. The results suggest that it is possible to reach a counting rate which is by a factor of eight higher in comparison to the rate of a detector with only one double-gap RPC. The effect of neutron scattering inside the detector contributing to the background is analyzed and design modifications of the first detector prototype, tested at neutron beam, are suggested.