Do you want to publish a course? Click here

In this experimental work, the instruments used (detector, calibrating sources) has been calibrating, the calibration line of the detector deduced and it has been finding that the equation of calibration line takes the form: The resolution power R has been studying as function of the distance between the activity source and detector; the results obtained show that R decrease by increase the energy of the lines spectrum, this mean that the detector resolution power become bigger. It has been measured the LC، LD and MDA using two different methods of background radiation value, trapezium method and classical method, for two different distances between radioactivity source and detector: (9,3cm) and (10cm). The results show: 1- There is similar comportment of the LC، LD and MDA in function energy and the distance between radioactivity source and detector. 2- The trapezium method gives better results from the classical method to the radiation background, and this leads to lesser MDA value, this means that the sensibility is bigger to measure this quantity. 3- The relative error using MDA is (5%) for large values, and (10%) for small values
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا