A major advance in accurate electron beam polarization measurement has been achieved at Jlab Hall A with a Compton polarimeter based on a Fabry-Perot cavity photon beam amplifier. At an electron energy of 4.6 GeV and a beam current of 40 uA, a total relative uncertainty of 1.5% is typically achieved within 40 min of data taking. Under the same conditions monitoring of the polarization is accurate at a level of 1%. These unprecedented results make Compton polarimetry an essential tool for modern parity-violation experiments, which require very accurate electron beam polarization measurements.
A wide range of nucleon and nuclear structure experiments in Jefferson Labs Hall A require precise, continuous measurements of the polarization of the electron beam. In our Compton polarimeter, electrons are scattered off photons in a Fabry-Perot cav
We report on the highest precision yet achieved in the measurement of the polarization of a low energy, $mathcal{O}$(1 GeV), electron beam, accomplished using a new polarimeter based on electron-photon scattering, in Hall~C at Jefferson Lab. A number
Compton backscattering polarimetry provides a fast and accurate method to measure the polarization of an electron beam in a storage ring. Since the method is non-destructive, the polarization of the electron beam can be monitored during internal targ
Compton backscattering polarimetry provides a fast measurement of the polarization of an electron beam in a storage ring. Since the method is non-destructive, the polarization of the electrons can be monitored during internal target experiments. At N
A lead-glass hodoscope calorimeter that was constructed for use in the Jefferson Lab Real Compton Scattering experiment is described. The detector provides a measurement of the coordinates and the energy of scattered photons in the GeV energy range w