Detection efficiency evaluation for a large area neutron sensitive microchannel plate detector

107   0   0.0 ( 0 )
 نشر من قبل Yiming Wang
 تاريخ النشر 2015
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

In this paper, the detection efficiency of a large area neutron sensitive microchannel plate detector has been evaluated. A 6LiF/ZnS detector was employed as the benchmark detector, the TOF spectra of these two detectors were simultaneously measured and the energy spectra were then deduced to calculate the detection efficiency curve of the nMCP detector. Tests show the detection [email protected] meV thermal neutron is 34% for this nMCP detector.

قيم البحث

اقرأ أيضاً

A beam imaging detector was developed by coupling a multi-strip anode with delay line readout to an E$times$B microchannel plate (MCP) detector. This detector is capable of measuring the incident position of the beam particles in one-dimension. To as sess the spatial resolution, the detector was illuminated by an $alpha$-source with an intervening mask that consists of a series of precisely-machined slits. The measured spatial resolution was 520$mu$m FWHM, which was improved to 413$mu$m FWHM by performing an FFT of the signals, rejecting spurious signals on the delay line, and requiring a minimum signal amplitude. This measured spatial resolution of 413$mu$m FWHM corresponds to an intrinsic resolution of 334$mu$m FWHM when the effect of the finite slit width is de-convoluted. To understand the measured resolution, the performance of the detector is simulated with the ion-trajectory code SIMION.
We previously proposed a method to detect neutrons by using a current-biased kinetic inductance detector (CB-KID), where neutrons are converted into charged particles using a 10B conversion layer. The charged particles are detected based on local cha nges in kinetic inductance of X and Y superconducting meanderlines under a modest DC bias current. The system uses a delay-line method to locate the positions of neutron-10B reactions by acquiring the four arrival timestamps of signals that propagate from hot spots created by a passing charged particle to the end electrodes of the meanderlines. Unlike conventional multi-pixel imaging systems, the CB-KID system performs high spatial resolution imaging over a 15 mm x 15 mm sensitive area using only four channel readouts. Given the large sensitive area, it is important to check the spatial homogeneity and linearity of detected neutron positions when imaging with CB-KID. To this end we imaged a pattern of 10B dot absorbers with a precise dot pitch to investigate the spatial homogeneity of the detector. We confirmed the spatial homogeneity of detected dot positions based on the distribution of measured dot pitches across the sensitive area of the detector. We demonstrate potential applications of the system by taking a clear transmission image of tiny metallic screws and nuts and a ladybug. The image was useful for characterizing the ladybug noninvasively. Detection efficiencies were low when the detector was operated at 4 K, so we plan to explore raising the operating temperature towards the critical temperature of the detector as a means to improve counting rates.
The neutrino detector of the Jiangmen Underground Neutrino Observatory (JUNO) is designed to use 20 kilotons of liquid scintillator and approximately 16,000 20-inch photomultipliers (PMTs).One of the options is to use the 20-inch R12860 PMT with high quantum efficiency which has recently been developed by Hamamatsu Photonics. The performance of the newly developed PMT preproduction samples is evaluated. The results show that its quantum efficiency is $30%$ at $400 nm$. Its Peak/Valley (P/V) ratio for the single photoelectron is 4.75 and the dark count rate is $27 kHz$ at the threshold of 3 mV while the gain is at $1 times 10^7$. The transit time spread of a single photoelectron is $2.86 ns$. Generally the performances of this new 20-inch PMT are improved over the old one of R3600.
An efficient neutron detection system with good energy resolution is required to correctly characterize decays of neutron-rich nuclei where $beta-$delayed neutron emission is a dominant decay mode. The Neutron dEtector with Xn Tracking (NEXT) has bee n designed to measure $beta$-delayed neutron emitters. By segmenting the detector along the neutron flight path, NEXT reduces the associated uncertainties in neutron time-of-flight measurements, improving energy resolution while maintaining detection efficiency. Detector prototypes are comprised of optically separated segments of a neutron-gamma discriminating plastic scintillator coupled to position-sensitive photomultiplier tubes. The first performance studies of this detector showed that high intrinsic neutron detection efficiency could be achieved while retaining good energy resolution. The results from the efficiency measurements using neutrons from direct reactions are presented.
We show data from a new type of detector that can be used to determine neutron flux, energy distribution, and direction of neutron motion for both fast and thermal neutrons. Many neutron detectors are plagued by large backgrounds from x-rays and gamm a rays, and most current neutron detectors lack single-event energy sensitivity or any information on neutron directionality. Even the best detectors are limited by cosmic ray neutron backgrounds. All applications (neutron scattering and radiography, measurements of solar and cosmic ray neutron flux, measurements of neutron interaction cross sections, monitoring of neutrons at nuclear facilities, oil exploration, and searches for fissile weapons of mass destruction) will benefit from the improved neutron detection sensitivity and improved measurements of neutron properties made possible by this detector. The detector is free of backgrounds from x-rays, gamma rays, beta particles, relativistic singely charged particles and cosmic ray neutrons. It is sensitive to thermal neutrons, fission neutrons, and high energy neutrons, with detection features distinctive for each energy range. It is capable of determining the location of a source of fission neutrons based on characteristics of elastic scattering of neutrons by helium nuclei. The detector we have constructed could identify one gram of reactor grade plutonium, one meter away, with less than one minute of observation time.
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا