The Flight Spectral Response of the ACIS Instrument

91   0   0.0 ( 0 )
 نشر من قبل Paul P. Plucinsky
 تاريخ النشر 2002
  مجال البحث فيزياء
والبحث باللغة English
 تأليف Paul P. Plucinsky




اسأل ChatGPT حول البحث

We discuss the flight calibration of the spectral response of the Advanced CCD Imaging Spectrometer (ACIS) on-board the Chandra X-ray Observatory (CXO). The spectral resolution and sensitivity of the ACIS instrument have both been evolving over the course of the mission. The spectral resolution of the frontside-illuminated (FI) CCDs changed dramatically in the first month of the mission due to radiation damage. Since that time, the spectral resolution of the FI CCDs and the backside-illuminated (BI) CCDs have evolved gradually with time. We demonstrate the efficacy of charge-transfer inefficiency (CTI) correction algorithms which recover some of the lost performance. The detection efficiency of the ACIS instrument has been declining throughout the mission, presumably due to a layer of contamination building up on the filter and/or CCDs. We present a characterization of the energy dependence of the excess absorption and demonstrate software which models the time dependence of the absorption from energies of 0.4 keV and up. The spectral redistribution function and the detection efficiency are well-characterized at energies from 1.5 to 8.0 keV. The calibration at energies below 1.5 keV is challenging because of the lack of strong lines in the calibration source and also because of the inherent non-linear dependence with energy of the CTI and the absorption by the contamination layer. We have been using data from celestial sources with relatively simple spectra to determine the quality of the calibration below 1.5 keV. The analysis of these observations demonstrate that the CTI correction recovers a significant fraction of the spectral resolution of the FI CCDs and the models of the time-dependent absorption result in consistent measurements of the flux at low energies for data from a BI (S3) CCD.



قيم البحث

اقرأ أيضاً

We present the first results of the in-flight validation of the spectral response of the FREGATE X/gamma detectors on-board the HETE-2 satellite. This validation uses the Crab pulsar and nebula as reference spectra.
SPIRE, the Spectral and Photometric Imaging Receiver, is the Herschel Space Observatorys submillimetre camera and spectrometer. It contains a three-band imaging photometer operating at 250, 350 and 500 {mu}m, and an imaging Fourier transform spectrom eter (FTS) covering 194-671 {mu}m (447-1550 GHz). In this paper we describe the initial approach taken to the absolute calibration of the SPIRE instrument using a combination of the emission from the Herschel telescope itself and the modelled continuum emission from solar system objects and other astronomical targets. We present the photometric, spectroscopic and spatial accuracy that is obtainable in data processed through the standard pipelines. The overall photometric accuracy at this stage of the mission is estimated as 15% for the photometer and between 15 and 50% for the spectrometer. However, there remain issues with the photometric accuracy of the spectra of low flux sources in the longest wavelength part of the SPIRE spectrometer band. The spectrometer wavelength accuracy is determined to be better than 1/10th of the line FWHM. The astrometric accuracy in SPIRE maps is found to be 2 arcsec when the latest calibration data are used. The photometric calibration of the SPIRE instrument is currently determined by a combination of uncertainties in the model spectra of the astronomical standards and the data processing methods employed for map and spectrum calibration. Improvements in processing techniques and a better understanding of the instrument performance will lead to the final calibration accuracy of SPIRE being determined only by uncertainties in the models of astronomical standards.
The Marshall Grazing Incidence X-ray Spectrometer (MaGIXS) is a sounding rocket experiment that observes the soft X-ray spectrum of the Sun from 6.0 - 24 Angstrom (0.5 - 2.0 keV), successfully launched on 30 July 2021. End-to-end alignment of the fli ght instrument and calibration experiments are carried out using the X-ray and Cryogenic Facility (XRCF) at NASA Marshall Space Flight Center. In this paper, we present the calibration experiments of MaGIXS, which include wavelength calibration, measurement of line spread function, and determination of effective area. Finally, we use the measured instrument response function to predict the expected count rates for MaGIXS flight observation looking at a typical solar active region
Since the MICROSCOPE instrument aims to measure accelerations as low as a few 10$^{-15}$,m,s$^{-2}$ and cannot operate on ground, it was obvious to have a large time dedicated to its characterization in flight. After its release and first operation, the characterization experiments covered all the aspects of the instrument design in order to consolidate the scientific measurements and the subsequent conclusions drawn from them. Over the course of the mission we validated the servo-control and even updated the PID control laws for each inertial sensor. Thanks to several dedicated experiments and the analysis of the instrument sensitivities, we have been able to identify a number of instrument characteristics such as biases, gold wire and electrostatic stiffnesses, non linearities, couplings and free motion ranges of the test-masses, which may first impact the scientific objective and secondly the analysis of the instrument good operation.
581 - O. Godet 2008
(Abbreviated) We show that the XRT spectral response calibration was complicated by various energy offsets in photon counting (PC) and windowed timing (WT) modes related to the way the CCD is operated in orbit (variation in temperature during observa tions, contamination by optical light from the sunlit Earth and increase in charge transfer inefficiency). We describe how these effects can be corrected for in the ground processing software. We show that the low-energy response, the redistribution in spectra of absorbed sources, and the modelling of the line profile have been significantly improved since launch by introducing empirical corrections in our code when it was not possible to use a physical description. We note that the increase in CTI became noticeable in June 2006 (i.e. 14 months after launch), but the evidence of a more serious degradation in spectroscopic performance (line broadening and change in the low-energy response) due to large charge traps (i.e. faults in the Si crystal) became more significant after March 2007. We describe efforts to handle such changes in the spectral response. Finally, we show that the commanded increase in the substrate voltage from 0 to 6V on 2007 August 30 reduced the dark current, enabling the collection of useful science data at higher CCD temperature (up to -50C). We also briefly describe the plan to recalibrate the XRT response files at this new voltage.
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا