ترغب بنشر مسار تعليمي؟ اضغط هنا

Ultra-clean radon-free four cylinder magnetically-coupled piston pump

181   0   0.0 ( 0 )
 نشر من قبل Denny Schulte
 تاريخ النشر 2021
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

A high performance gas displacement pump based on four individual cylinders was developed. The magnetically-coupled pistons and the hermetically-sealed housings as well as the special cleanliness in terms of out-gassing and radio-purity make the four cylinder pump interesting for the usage in low background experiments dealing with noble gases as target. An optimized inner flow path as well as reduced dead volume with respect to the prototype single cylinder pump of the XENON1T experiment combined with advanced polymer piston gaskets and a piston coupling force of (3177 $pm$ 162) N ensures a high and reliable performance. Furthermore, a phase-shifted synchronization of the pistons movement creates an additional performance boost as well as minimizes fluctuations of flow and compression. The pump is able to generate xenon flows of 474 slpm at a pressure difference of 1.80 bar.

قيم البحث

اقرأ أيضاً

A magnetically driven piston pump for xenon gas recirculation is presented. The pump is designed to satisfy extreme purity and containment requirements, as is appropriate for the recirculation of isotopically enriched xenon through the purification s ystem and large liquid xenon TPC of EXO-200. The pump, using sprung polymer gaskets, is capable of pumping more than 16 standard liters per minute (SLPM) of xenon gas with 750 torr differential pressure.
The MAJORANA DEMONSTRATOR is a neutrinoless double beta decay experiment utilizing enriched Ge-76 detectors in 2 separate modules inside of a common solid shield at the Sanford Underground Research Facility. The DEMONSTRATOR has utilized world leadin g assay sensitivities to develop clean materials and processes for producing ultra-pure copper and plastic components. This experiment is now operating, and initial data provide new insights into the success of cleaning and processing. Post production copper assays after the completion of Module 1 showed an increase in U and Th contamination in finished parts compared to starting bulk material. A revised cleaning method and additional round of surface contamination studies prior to Module 2 construction have provided evidence that more rigorous process control can reduce surface contamination. This article describes the assay results and discuss further studies to take advantage of assay capabilities for the purpose of maintaining ultra clean fabrication and process design.
We describe a compact, ultra-clean device used to deploy radioactive sources along the vertical axis of the KamLAND liquid-scintillator neutrino detector for purposes of calibration. The device worked by paying out and reeling in precise lengths of a hanging, small-gauge wire rope (cable); an assortment of interchangeable radioactive sources could be attached to a weight at the end of the cable. All components exposed to the radiopure liquid scintillator were made of chemically compatible UHV-cleaned materials, primarily stainless steel, in order to avoid contaminating or degrading the scintillator. To prevent radon intrusion, the apparatus was enclosed in a hermetically sealed housing inside a glove box, and both volumes were regularly flushed with purified nitrogen gas. An infrared camera attached to the side of the housing permitted real-time visual monitoring of the cables motion, and the system was controlled via a graphical user interface.
The LZ dark matter detector, like many other rare-event searches, will suffer from backgrounds due to the radioactive decay of radon daughters. In order to achieve its science goals, the concentration of radon within the xenon should not exceed $2mu$ Bq/kg, or 20 mBq total within its 10 tonnes. The LZ collaboration is in the midst of a program to screen all significant components in contact with the xenon. The four institutions involved in this effort have begun sharing two cross-calibration sources to ensure consistent measurement results across multiple distinct devices. We present here five preliminary screening results, some mitigation strategies that will reduce the amount of radon produced by the most problematic components, and a summary of the current estimate of radon emanation throughout the detector. This best estimate totals $<17.3$ mBq, sufficiently low to meet the detectors science goals.
The decay of common radioactive gases, such as radon, produces stable isotopes by a sequence of daughter particles with varied half-lives. These daughter particles are a significant source of gamma, neutron, and alpha particle backgrounds that can mi mic desired signals in dark matter and neutrinoless double beta decay experiments. In the LUMINA Laboratory at Southern Methodist University (SMU), studies of radon plate-out onto copper samples are conducted using one of XIAs first five UltraLo 1800 alpha counters. We present results from investigations into various mitigation approaches. A custom-built copper holder (in either plastic or metal) has been designed and produced to maximize the coppers exposure to 220Rn. The 220Rn source is a collection of camping lantern mantles. We present the current status of control and experimental methods for addressing radon exposure levels.