ترغب بنشر مسار تعليمي؟ اضغط هنا

The Cesium Source Calibration and Monitoring System of the ATLAS Tile Calorimeter: Design, Construction and Results

88   0   0.0 ( 0 )
 نشر من قبل Oleg Solovyanov
 تاريخ النشر 2020
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

This article describes the design, construction and use of a calibration and monitoring system, based on movable 137Cs gamma-ray sources, for the ATLAS Tile Calorimeter (TileCal). The sources, propelled by a water-based liquid through tubes that traverse all the calorimeters cells, produce signals that precisely characterise the response of each tile, thereby providing very granular and accurate data on the response of TileCal to particles. The system has been used to guide and control the quality of the optical instrumentation of all TileCal modules, to set and equalise the dynamic range of the response to physics data, and to set the energy scale of the readout system. In the ATLAS cavern, periodic measurements of the whole detectors response to 137Cs sources allow monitoring the uniformity and stability of all the calorimeters cells as well as maintaining precise knowledge of its energy calibration. The design of the source hydraulic drive systems hardware and software, the data acquisition system and the data processing algorithms are described. Finally, the results of this two-decade program are shown.

قيم البحث

اقرأ أيضاً

88 - D. Calvet 2018
This article documents the characteristics of the high voltage (HV) system of the hadronic calorimeter TileCal of the ATLAS experiment. Such a system is suitable to supply reliable power distribution into particles physics detectors using a large num ber of PhotoMultiplier Tubes (PMTs). Measurements performed during the 2015 and 2016 data taking periods of the ATLAS detector show that its performance, in terms of stability and noise, fits the specifications. In particular, almost all the PMTs show a voltage instability smaller than 0.5 V corresponding to a gain stability better than 0.5%. A small amount of channels was found not working correctly. To diagnose the origin of such defects, the results of the HV measurements were compared to those obtained using a Laser system. The analysis shows that less than 0.2% of the about 10 thousand HV channels were malfunctioning.
87 - T. Barillari 2004
ATLAS has chosen for its Hadronic End-Cap Calorimeter (HEC) the copper-liquid argon sampling technique with flat plate geometry and GaAs pre-amplifiers in the argon. The contruction of the calorimeter is now approaching completion. Results of product ion quality checks are reported and their anticipated impact on calorimeter performance discussed. Selected results, such as linearity, electron and pion energy resolution, uniformity of energy response, obtained in beam tests both of the Hadronic End-Cap Calorimeter by itself, and in the ATLAS configuration where the HEC is in combination with the Electromagnetic End-Cap Calorimeter (EMEC) are described.
118 - F. Ghio , B. Girolami , M. Capogni 1997
We describe the electromagnetic calorimeter built for the GRAAL apparatus at the ESRF. Its monitoring system is presented in detail. Results from tests and the performance obtained during the first GRAAL experiments are given. The energy calibration accuracy and stability reached is a small fraction of the intrinsic detector resolution.
90 - Jalal Abdallah 2020
The ATLAS hadronic Tile Calorimeter will undergo major upgrades to the on- and off-detector electronics in preparation for the High Luminosity program of the Large Hadron Collider (HL-LHC) in 2026, so that the system can cope with the HL-LHC increase d radiation levels and out-of-time pileup. The on-detector electronics of the upgraded system will continuously digitize and transmit all photo-multiplier signals to the off-detector systems at a 40 MHz rate. The off-detector electronics will store the data in pipeline buffers, produce digital hadronic tower sums for the ATLAS Level-0 trigger system, and read out selected events. The modular on-detector electronics feature radiation-tolerant commercial off-the-shelf components and redundant design to minimize single points of failure. The timing, control and communication interface with the off-detector electronics is implemented with modern Field Programmable Gate Arrays and high speed fibre optic links running up to 9.6 Gbps.
84 - C. Adams , A. Bambaugh , B. Bilki 2016
A novel hadron calorimeter is being developed for future lepton colliding beam detectors. The calorimeter is optimized for the application of Particle Flow Algorithms (PFAs) to the measurement of hadronic jets and features a very finely segmented rea dout with 1 x 1 cm2 cells. The active media of the calorimeter are Resistive Plate Chambers (RPCs) with a digital, i.e. one-bit, readout. To first order the energy of incident particles in this calorimeter is reconstructed as being proportional to the number of pads with a signal over a given threshold. A large-scale prototype calorimeter with approximately 500,000 readout channels has been built and underwent extensive testing in the Fermilab and CERN test beams. This paper reports on the design, construction, and commissioning of this prototype calorimeter.
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا