A nanometric needle sensor mounted in an Atomic Force Microscopy allows systematic picometer-range distance measurements. This force sensing device is used in Piezoresponse Force Microscopy (PFM) as a distance sensor, by employing the cantilever spring constant as the conversion factor opening a pathway to explore the piezoelectric effect at the nanoscale. The force-distance equivalence is achieved if the force sensor does not disturb the system to study, solely. In this manuscript we report a systematic study in which different Lead Zirconate Titanate (PZT) materials, having different d33 values, are measured following the standard theory available for PFM. Both in resonance and out of resonance measurements demonstrate that PFM cannot be considered quantitative. After performing the measurements, we propose a correction of the standard theory employed in PFM by considering the force exerted by the material as a variable. The g33 parameter, inherent to piezoelectricity, governs the amount of force available from the system. A comparison of piezoelectric stiffness for the case of a nanoscale site contact region, similar to the one it is found while performing PFM, is provided. Such stiffness is well below the cantilever stiffness, limiting and diminishing the material movement, as the piezoelectric material does not have enough stroke to induce the intended displacement. A correction factor, named Open Piezopotential Gauge, accounts for these effects, which is used to correct the measurements carried out in PZT materials towards a real quantitative PFM.

قيم البحث

اقرأ أيضاً

To achieve quantitative interpretation of Piezoresponse Force Microscopy (PFM), including resolution limits, tip bias- and strain-induced phenomena and spectroscopy, analytical representations for tip-induced electroelastic fields inside the material are derived for the cases of weak and strong indentation. In the weak indentation case, electrostatic field distribution is calculated using image charge model. In the strong indentation case, the solution of the coupled electroelastic problem for piezoelectric indentation is used to obtain the electric field and strain distribution in the ferroelectric material. This establishes a complete continuum mechanics description of the PFM contact mechanics and imaging mechanism. The electroelastic field distribution allows signal generation volume in PFM to be determined. These rigorous solutions are compared with the electrostatic point charge and sphere-plane models, and the applicability limits for asymptotic point charge and point force models are established. The implications of these results for ferroelectric polarization switching processes are analyzed.
The nondestructive imaging of subsurface structures on the nanometer scale has been a long-standing desire in both science and industry. A few impressive images were published so far that demonstrate the general feasibility by combining ultrasound wi th an Atomic Force Microscope. From different excitation schemes, Heterodyne Force Microscopy seems to be the most promising candidate delivering the highest contrast and resolution. However, the physical contrast mechanism is unknown, thereby preventing any quantitative analysis of samples. Here we show that friction at material boundaries within the sample is responsible for the contrast formation. This result is obtained by performing a full quantitative analysis, in which we compare our experimentally observed contrasts with simulations and calculations. Surprisingly, we can rule out all other generally believed responsible mechanisms, like Rayleigh scattering, sample (visco)elasticity, damping of the ultrasonic tip motion, and ultrasound attenuation. Our analytical description paves the way for quantitative SubSurface-AFM imaging.
Frequency dependent dynamic behavior in Piezoresponse Force Microscopy (PFM) implemented on a beam-deflection atomic force microscope (AFM) is analyzed using a combination of modeling and experimental measurements. The PFM signal comprises contributi ons from local electrostatic forces acting on the tip, distributed forces acting on the cantilever, and three components of the electromechanical response vector. These interactions result in the bending and torsion of the cantilever, detected as vertical and lateral PFM signals. The relative magnitudes of these contributions depend on geometric parameters of the system, the stiffness and frictional forces of tip-surface junction, and operation frequencies. The dynamic signal formation mechanism in PFM is analyzed and conditions for optimal PFM imaging are formulated. The experimental approach for probing cantilever dynamics using frequency-bias spectroscopy and deconvolution of electromechanical and electrostatic contrast is implemented.
Domain dynamics in the Piezoresponse Force Spectroscopy (PFS) experiment is studied using the combination of local hysteresis loop acquisition with simultaneous domain imaging. The analytical theory for PFS signal from domain of arbitrary cross-secti on is developed and used for the analysis of experimental data on Pb(Zr,Ti)O3 polycrystalline films. The results suggest formation of oblate domain at early stage of the domain nucleation and growth, consistent with efficient screening of depolarization field within the material. The fine structure of the hysteresis loop is shown to be related to the observed jumps in the domain geometry during domain wall propagation (nanoscale Barkhausen jumps), indicative of strong domain-defect interactions.
Nanoscale polarization switching in ferroelectric materials by Piezoresponse Force Microscopy (PFM) in weak and strong indentation limits is analyzed using exact solutions for electrostatic and coupled electroelastic fields below the tip. It is propo sed that the tip-induced domain switching can be mapped on the Landau theory of phase transitions with the domain size as an order parameter. For a point charge interacting with a ferroelectric surface, switching of both first and second order is possible depending on the charge-surface separation. For a realistic tip shape, the domain nucleation process is first order in charge magnitude and polarization switching occurs only above a critical tip bias. In pure ferroelectric or ferroelastic switching, the late stages of the switching process can be described using point charge/force model and arbitrarily large domains can be created; however, the description of the early stages of nucleation process when domain size is comparable with the tip radius of curvature requires exact field structure to be taken into account.