ﻻ يوجد ملخص باللغة العربية
We measure the current noise of several cryogenic cables in a pulse tube based dilution refrigerator at frequencies between about 1~mHz and 50~kHz. We show that vibration-induced noise can be efficiently suppressed by using vacuum-insulated cables between room temperature and the 2nd pulse tube stage. A noise peak below 4 fA at the 1.4~Hz operation frequency of the pulse tube, and a white noise density of 0.44 fA/sqrt{Hz} in the millihertz range are obtained.
In this work we present data characterizing the sensitivity of the Bir{e}fringence Magnetique du Vide (BMV) instrument. BMV is an experiment attempting to measure vacuum magnetic birefringence (VMB) via the measurement of an ellipticity induced in a
We implemented a low noise current preamplifier for the readout of resistive bolometers. We tested the apparatus on thermometer resistances ranging from 10 Ohm to 500 Mohm. The use of current preamplifier overcomes constraints introduced by the reado
In this contribution we review the status and perspectives of direct neutrino mass experiments. These experiments investigate the kinematics of $beta$-decays of specific isotopes ($^3$H, $^{187}$Re, $^{163}$Ho) to derive model-independent information
We report a homebuilt ultra-high-vacuum (UHV) rotating sample manipulator with cryogenic cooling. The sample holder is thermally anchored to a built-in cryogenic cold head through flexible copper beryllium strips, permitting continuous sample rotatio
In this paper we describe the technology of building a vacuum-tight high voltage feedthrough which is able to operate at voltages up to 30 kV. The feedthrough has a coaxial structure with a grounded sheath which makes it capable to lead high voltage