Code-division multiplexing for x-ray microcalorimeters

506   0   0.0 ( 0 )
 نشر من قبل Joseph Fowler
 تاريخ النشر 2012
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

We demonstrate the code-division multiplexed (CDM) readout of eight transition-edge sensor microcalorimeters. The energy resolution is 3.0 eV (full width at half-maximum) or better at 5.9 keV, with a best resolution of 2.3 eV and a mean of 2.6 eV over the seven modulated detectors. The flux-summing CDM system is described and compared with similar time-division multiplexed (TDM) readout. We show that the sqrt(Npixels) multiplexing disadvantage associated with TDM is not present in CDM. This demonstration establishes CDM as both a simple route to higher performance in existing TDM microcalorimetric experiments and a long-term approach to reaching higher multiplexing factors.



قيم البحث

اقرأ أيضاً

We are developing code-division multiplexing (CDM) systems for transition-edge sensor arrays with the goal of reaching multiplexing factors in the hundreds. We report on x-ray measurements made with a four-channel prototype CDM system that employs a flux-summing architecture, emphasizing data-analysis issues. We describe an empirical method to determine the demodulation matrix that minimizes cross-talk. This CDM system achieves energy resolutions of between 2.3 eV and 3.0 eV FWHM at 5.9 keV.
The Transition-Edge Sensor (TES) is an extremely sensitive device which is used to measure the energy of individual X-ray photons. For astronomical spectrometry applications, SRON develops a Frequency Domain Multiplexing (FDM) read-out system for kil opixel arrays of such TESs. Each TES is voltage biased at a specific frequency in the range 1 to 5 MHz. Isolation between the individual pixels is obtained through very narrow-band (high-Q) lithographic LC resonators. To prevent energy resolution degradation due to intermodulation line noise, the bias frequencies are distributed on a regular grid. The requirements on the accuracy of the LC resonance frequency are very high. The deviation of the resonance frequencies due to production tolerances is significant with respect to the bandwidth, and a controller is necessary to compensate for the LC series impedance. We present two such controllers: a simple orthogonal proportional-integrating (PI) controller and a more complex impedance estimator. Both controllers operate in baseband and try to make the TES current in-phase with the bias voltage, effectively operating as phase-locked loops (PLL). They allow off-LC-resonance operation of the TES pixels, while preserving TES thermal response and energy resolution. Extensive experimental results -- published in a companion paper recently -- with the proposed methods, show that these controllers allow the preservation of single pixel energy resolution in multiplexed operation.
We are developing X-ray microcalorimeters as a backup option for the baseline detectors in the X-IFU instrument on board the ATHENA space mission led by ESA and to be launched in the early 2030s.5$times$5 mixed arrays with TiAu transition-edge sensor (TES), which have different high aspect ratios and thus high resistances, have been designed and fabricated to meet the energy resolution requirement of the X-IFU instrument. Such arrays can also be used to optimise the performance of the Frequency Domain Multiplexing (FDM) readout and lead to the final steps for the fabrication of a large detector array. In this work we present the experimental results from tens of the devices with an aspect ratio (length-to-width) ranging from 1-to-1 up to 6-to-1, measured in a single-pixel mode with a FDM readout system developed at SRON/VTT. We observed a nominal energy resolution of about 2.5 eV at 5.9 keV at bias frequencies ranging from 1 to 5 MHz. These detectors are proving to be the best TES microcalorimeters ever reported in Europe, being able to meet not only the requirements of the X-IFU instrument, but also those of other future challenging X-ray space missions, fundamental physics experiments, plasma characterization and material analysis.
Transition-edge sensor X-ray microcalorimeters are usually calibrated empirically, as the most widely-used calibration metric, optimal filtered pulse height (OFPH), in general has an unknown dependance on photon energy, $E_{gamma}$. Because the calib ration function can only be measured at specific points where photons of a known energy can be produced, this unknown dependence of OFPH on $E_{gamma}$ leads to calibration errors and the need for time-intensive calibration measurements and analysis. A calibration metric that is nearly linear as a function of $E_{gamma}$ could help alleviate these problems. In this work, we assess the linearity of a physically motivated calibration metric, $E_{Joule}$. We measure calibration pulses in the range 4.5 keV$<$$E_{gamma}$$<$9.6 keV with detectors optimized for 6 keV photons to compare the linearity properties of $E_{Joule}$ to OFPH. In these test data sets, we find that $E_{Joule}$ fits a linear function an order of magnitude better than OFPH. Furthermore, calibration functions using $E_{J}$, an optimized version of $E_{Joule}$, are linear within the 2-3 eV noise of the data.
The determination of the neutrino mass is an open issue in modern particle physics and astrophysics. The direct mass measurement is the only theory-unrelated experimental tool capable to probe such quantity. The HOLMES experiment aims to measure the end-point energy of the electron capture (EC) decay of $^{163}$Ho with a statistical sensitivity on the neutrino mass as low as $sim 1$ eV/c$^2$. In order to acquire the large needed statistics, by keeping the pile-up contribution as low as possible, 1024 transition edge sensors (TESs) with high energy and time resolutions will be employed. Microcalorimeter and bolometer arrays based on transition edge sensor with thousands of pixels are under development for several space-based and ground-based applications, including astrophysics, nuclear and particle physics, and materials science. The common necessary challenge is to develop pratical multiplexing techniques in order to simplify the cryogenics and readout systems. Despite the various multiplexing variants which are being developed have been successful, new approaches are needed to enable scaling to larger pixel counts and faster sensors, as requested for HOLMES, reducing also the cost and complexity of readout. A very novel technique that meets all of these requirements is based on superconducting microwave resonators coupled to radio-frequency Superconducting Quantum Interference Devices, in which the the changes in the TES input current is tranduced to a change in phase of a microwave signal. In this work we introduce the basics of this technique, the design and development of the first two-channel read out system and its performances with the first TES detectors specifically designed for HOLMES. In the last part we explain how to extend this approach scaling to 1024 pixels.