Performance of Multi-Pixel Photon Counters for the T2K near detectors

177   0   0.0 ( 0 )
 نشر من قبل Masashi Yokoyama
 تاريخ النشر 2010
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

We have developed a Multi-Pixel Photon Counter (MPPC) for the neutrino detectors of T2K experiment. About 64,000 MPPCs have been produced and tested in about a year. In order to characterize a large number of MPPCs, we have developed a system that simultaneously measures 64 MPPCs with various bias voltage and temperature. The performance of MPPCs are found to satisfy the requirement of T2K experiment. In this paper, we present the performance of 17,686 MPPCs measured at Kyoto University.

قيم البحث

اقرأ أيضاً

137 - G. Blazey 2019
Results of radiation tests of Hamamatsu 2.0 x 2.0~mm2 through-silicon-via (S13360-2050VE) multi-pixel photon counters, or MPPCs [1], are presented. Distinct sets of eight MPPCs were exposed to four different 1~MeV neutron equivalent doses of 200 MeV protons. Measurements of the breakdown voltage, gain and noise rates at different bias overvoltages, photoelectron thresholds, and LED illumination levels were taken before and after irradiation. No significant deterioration in performance was observed for breakdown voltage, gain, and response. Noise rates increased significantly with irradiation. These studies were undertaken in the context of MPPC requirements for the Cosmic Ray Veto detector of the Mu2e experiment at the Fermi National Accelerator Laboratory.
The multi-pixel photon counter (MPPC) is a newly developed photodetector with an excellent photon counting capability. It also has many attractive features such as small size, high gain, low operation voltage and power consumption, and capability of operating in magnetic fields and in room temperature. The basic performance of samples has been measured. A gain of ~10^6 is achieved with a noise rate less than 1 MHz with 1 p.e. threshold, and cross-talk probability of less than 30% at room temperature. The photon detection efficiency for green light is twice or more that of the photomultiplier tubes. It is found that the basic performance of the MPPC is satisfactory for use in real experiments.
259 - A. Vacheret 2011
The calorimeter, range detector and active target elements of the T2K near detectors rely on the Hamamatsu Photonics Multi-Pixel Photon Counters (MPPCs) to detect scintillation light produced by charged particles. Detailed measurements of the MPPC ga in, afterpulsing, crosstalk, dark noise, and photon detection efficiency for low light levels are reported. In order to account for the impact of the MPPC behavior on T2K physics observables, a simulation program has been developed based on these measurements. The simulation is used to predict the energy resolution of the detector.
After pioneering gaseous detectors of single photon for RICH applications using CsI solid state photocathodes in MWPCs within the RD26 collaboration and by the constructions for the RICH detector of the COMPASS experiment at CERN SPS, in 2016 we have upgraded COMPASS RICH by novel gaseous photon detectors based on MPGD technology. Four novel photon detectors, covering a total active area of 1.5~m$^2$, have been installed in order to cope with the challenging efficiency and stability requirements of the COMPASS physics programme. They are the first application in an experiment of MPGD-based single photon detectors. All aspects of the upgrade are presented, including engineering, mass production, quality assessment and performance. Perspectives for further developments in the field of gaseous single photon detectors are also indicated.
The second phase of the T2K experiment is expected to start data taking in autumn 2022. An upgrade of the Near Detector (ND280) is under development and includes the construction of two new Time Projection Chambers called High-Angle TPC (HA-TPC). The two endplates of these TPCs will be paved with eight Micromegas type charge readout modules. The Micromegas detector charge amplification structure uses a resistive anode to spread the charges over several pads to improve the space point resolution. This innovative technique is combined with the bulk-Micromegas technology to compose the Encapsulated Resistive Anode Micromegas detector. A prototype has been designed, built and exposed to an electron beam at the DESY II test beam facility. The data have been used to characterize the charge spreading and to produce a RC map. Spatial resolution better than 600 $mu$m and energy resolution better than 9% are obtained for all incident angles. These performances fulfil the requirements for the upgrade of the ND280 TPC.