213 - Ushasi Datta 2015
A prototype of Multi-strip Multi-gap Resistive Plate chamber (MMRPC) with active area 40 cm $times$ 20 cm has been developed at SINP, Kolkata. Detailed response of the developed detector was studied with the pulsed electron beam from ELBE at Helmholt z-Zentrum Dresden-Rossendorf. In this report the response of SINP developed MMRPC with different controlling parameters is described in details. The obtained time resolution ($sigma_t$) of the detector after slew correction was 91.5$ pm $3 ps. Position resolution measured along ($sigma_x$) and across ($sigma_y$) the strip was 2.8$pm$0.6 cm and 0.58 cm, respectively. The measured absolute efficiency of the detector for minimum ionizing particle like electron was 95.8$pm$1.3 $%$. Better timing resolution of the detector can be achieved by restricting the events to a single strip. The response of the detector was mainly in avalanche mode but a few percentage of streamer mode response was also observed. A comparison of the response of these two modes with trigger rate was studied