Current $gamma$-ray telescopes suffer from a gap in sensitivity in the energy range between 100keV and 100MeV, and no polarisation measurement has ever been done on cosmic sources above 1MeV. Past and present e$^+$e$^-$ pair telescopes are limited at lower energies by the multiple scattering of electrons in passive tungsten converter plates. This results in low angular resolution, and, consequently, a drop in sensitivity to point sources below 1GeV. The polarisation information, which is carried by the azimuthal angle of the conversion plane, is lost for the same reasons. HARPO (Hermetic ARgon POlarimeter) is an R&D program to characterise the operation of a gaseous detector (a Time Projection Chamber or TPC) as a high angular-resolution and sensitivity telescope and polarimeter for $gamma$ rays from cosmic sources. It represents a first step towards a future space instrument in the MeV-GeV range. We built and characterised a 30cm cubic demonstrator [SPIE 91441M], and put it in a polarised $gamma$-ray beam at the NewSUBARU accelerator in Japan. Data were taken at photon energies from 1.74MeV to 74MeV, and with different polarisation configurations. We describe the experimental setup in beam. We then describe the software we developed to reconstruct the photon conversion events, with special focus on low energies. We also describe the thorough simulation of the detector used to compare results. Finally we will present the performance of the detector as extracted from this analysis and preliminary measurements of the polarisation asymmetry. This beam-test qualification of a gas TPC prototype in a $gamma$-ray beam could open the way to high-performance $gamma$-ray astronomy and polarimetry in the MeV-GeV energy range in the near future.
Camera image sensors can be used to detect ionizing radiation in addition to optical photons. In particular, cosmic-ray muons are detected as long, straight tracks passing through multiple pixels. The distribution of track lengths can be related to the thickness of the active (depleted) region of the camera image sensor through the known angular distribution of muons at sea level. We use a sample of cosmic-ray muon tracks recorded by the Distributed Electronic Cosmic-ray Observatory to measure the thickness of the depletion region of the camera image sensor in a commercial smart phone, the HTC Wildfire S. The track length distribution prefers a cosmic-ray muon angular distribution over an isotropic distribution. Allowing either distribution, we measure the depletion thickness to be between 13.9~$mu$m and 27.7~$mu$m. The same method can be applied to additional models of image sensor. Once measured, the thickness can be used to convert track length to incident polar angle on a per-event basis. Combined with a determination of the incident azimuthal angle directly from the track orientation in the sensor plane, this enables direction reconstruction of individual cosmic-ray events.
Since the discovery of nuclear gamma-rays, its imaging has been limited to pseudo imaging, such as Compton Camera (CC) and coded mask. Pseudo imaging does not keep physical information (intensity, or brightness in Optics) along a ray, and thus is capable of no more than qualitative imaging of bright objects. To attain quantitative imaging, cameras that realize geometrical optics is essential, which would be, for nuclear MeV gammas, only possible via complete reconstruction of the Compton process. Recently we have revealed that Electron Tracking Compton Camera (ETCC) provides a well-defined Point Spread Function (PSF). The information of an incoming gamma is kept along a ray with the PSF and that is equivalent to geometrical optics. Here we present an imaging-spectroscopic measurement with the ETCC. Our results highlight the intrinsic difficulty with CCs in performing accurate imaging, and show that the ETCC surmounts this problem. The imaging capability also helps the ETCC suppress the noise level dramatically by ~3 orders of magnitude without a shielding structure. Furthermore, full reconstruction of Compton process with the ETCC provides spectra free of Compton edges. These results mark the first proper imaging of nuclear gammas based on the genuine geometrical optics.
We report the measurement of the emission time profile of scintillation from gamma-ray induced events in the XMASS-I 832 kg liquid xenon scintillation detector. Decay time constant was derived from a comparison of scintillation photon timing distributions between the observed data and simulated samples in order to take into account optical processes such as absorption and scattering in liquid xenon. Calibration data of radioactive sources, $^{55}$Fe, $^{241}$Am, and $^{57}$Co were used to obtain the decay time constant. Assuming two decay components, $tau_1$ and $tau_2$, the decay time constant $tau_2$ increased from 27.9 ns to 37.0 ns as the gamma-ray energy increased from 5.9 keV to 122 keV. The accuracy of the measurement was better than 1.5 ns at all energy levels. A fast decay component with $tau_1 sim 2$ ns was necessary to reproduce data. Energy dependencies of $tau_2$ and the fraction of the fast decay component were studied as a function of the kinetic energy of electrons induced by gamma-rays. The obtained data almost reproduced previously reported results and extended them to the lower energy region relevant to direct dark matter searches.
We report on the construction, tests, calibrations and commissioning of an Optical Readout Time Projection Chamber (O-TPC) detector operating with a CO2(80%) + N2(20%) gas mixture at 100 and 150 Torr. It was designed to measure the cross sections of several key nuclear reactions involved in stellar evolution. In particular, a study of the rate of formation of oxygen and carbon during the process of helium burning will be performed by exposing the chamber gas to intense nearly mono-energetic gamma-ray beams at the High Intensity Gamma Source (HIgS) facility. The O-TPC has a sensitive target-drift volume of 30x30x21 cm^3. Ionization electrons drift towards a double parallel grid avalanche multiplier, yielding charge multiplication and light emission. Avalanche induced photons from N2 emission are collected, intensified and recorded with a Charge Coupled Device (CCD) camera, providing two-dimensional track images. The events time projection (third coordinate) and the deposited energy are recorded by photomultipliers and by the TPC charge-signal, respectively. A dedicated VME-based data acquisition system and associated data analysis tools were developed to record and analyze these data. The O-TPC has been tested and calibrated with 3.183 MeV alpha-particles emitted by a 148Gd source placed within its volume with a measured energy resolution of 3.0%. Tracks of alpha and 12C particles from the dissociation of 16O and of three alpha-particles from the dissociation of 12C have been measured during initial in-beam test experiments performed at the HIgS facility at Duke University. The full detection system and its performance are described and the results of the preliminary in-beam test experiments are reported.