Design study of a scintronic crystal targeting tens of picoseconds time resolution for gamma ray imaging: the ClearMind detector


الملخص بالإنكليزية

We describe the concept of a new gamma ray scintronic detector targeting a time resolution of the order of 25 ps FWHM, with millimetric volume reconstruction and high detection efficiency. Its design consists of a monolithic large PbWO4 scintillating crystal with an efficient photocathode directly deposited on it. With an index of refraction higher for the photocathode than for the crystal, this design negates the total reflection effect of optical photons at the crystal/photo-detector optical interface, and thus largely improves optical coupling between the crystal and the photodetector. This allows to detect efficiently the Cherenkov light produced by 511 keV photoelectric