Measurement of the Scintillation Efficiency of Na Recoils in NaI(Tl) down to 10 keV Nuclear Recoil Energy relevant to Dark Matter Searches


الملخص بالإنكليزية

We present preliminary results of measurements of the quenching factor for Na recoils in NaI(Tl) at room temperature, made at a dedicated neutron facility at the University of Sheffield. Measurements have been performed with a 2.45 MeV mono-energetic neutron generator in the energy range from 10 keV to 100 keV nuclear recoil energy. A BC501A liquid scintillator detector was used to tag neutrons. Cuts on pulse-shape discrimination from the BC501A liquid scintillator detector and neutron time-of-flight were performed on pulses recorded by a digitizer with a 2 ns sampling time. Measured quenching factors range from 19% to 26%, in agreement with other experiments. From pulse-shape analysis, a mean time of pulses from electron and nuclear recoils are compared down to 2 keV electron equivalent energy.

تحميل البحث