The purpose of the present work is the study of reconstruction properties of a new Molecular Breast Imaging (MBI) device for the early diagnosis of breast cancer, in Limited Angle Tomography (LAT), by using two asymmetric detector heads with different collimators. The detectors face each other in anti-parallel viewing direction and, mild-compressing the breast phantom, they are able to reconstruct the inner tumour of the phantoms with only a limited number of projections using a dedicated maximum-likelihood expectation maximization (ML-EM) algorithm. Phantoms, MBI system, as well as Monte Carlo simulator using Geant 4 Application for Tomographic Emission (GATE) software, are briefly described. MBI systems model has been implemented in IDL (Interactive Data Visualization), in order to evaluate the best LAT configuration of the system and its reconstruction ability by varying tumours size, depth and uptake. LAT setup in real and simulated configurations, as well as the ML-EM method and the preliminary reconstruction results, are discussed.