Comprehensive procedure for personalized dosimetry in computed tomography


الملخص بالإنكليزية

The purpose of this work is to develop viable procedures for verifying the applicability of personalized dosimetry in computed tomography (CT) using Monte Carlo-based simulations. Mobile equipment together with customized software was developed and used for rapid, non-invasive determination of equivalent source models of CT scanners under clinical conditions. Standard and anthropomorphic CT dose phantoms equipped with real-time CT dose probes at five representative positions were scanned. The accumulated dose was measured during the scan at the five positions. ImpactMC, a Monte Carlo-based CT dose software program, was used to simulate the scan. The necessary inputs were obtained from the scan parameters, from the equivalent source models and from the material-segmented CT images of the phantoms. Post-scan 3D dose distributions in the phantoms were simulated and the dose values calculated at the five positions inside the phantom were compared to measured dose values. Initial results were obtained by means of a General Electric Optima CT 660 and a Toshiba (Canon) Aquilion ONE. In general, the measured and calculated dose values were within relative uncertainties that had been estimated to be less than 10%. The procedures developed, which allow the post-CT scan dose to be measured and calculated at five points inside anthropomorphic phantoms, were found to be viable and rapid. The procedures are applicable to any scanner type under clinical conditions. Results show that the procedures are well suited for verifying the applicability of personalized CT dosimetry based on post-scan Monte Carlo calculations.

تحميل البحث