A new prototype of STAR inner Time Projection Chamber (iTPC) MWPC sector has been fabricated and tested in an X-ray test system. The wire chamber built at Shandong University has a wire tension precision better than 6$%$ and wire pitch precision better than 10 $mu$m. The gas gain uniformity and energy resolution are measured to be better than 1$%$ (RMS) and 20$%$ (FWHM), respectively, using an $^{55}$Fe X-ray source. The iTPC upgrade project is to replace all 24 STAR TPC inner sectors as a crucial detector upgrade for the RHIC beam energy scan phase II program. The test results show that the constructed iTPC prototype meets all project requirements.