Relationship between mass density, electron density, and elemental composition of body tissues for Monte Carlo simulation in radiation treatment planning


الملخص بالإنكليزية

Purpose: For Monte Carlo simulation of radiotherapy, x-ray CT number of every system needs to be calibrated and converted to mass density and elemental composition. This study aims to formulate material properties of body tissues for practical two-step conversion from CT number. Methods: We used the latest compilation on body tissues that constitute reference adult male and female. We formulated the relations among mass, electron, and elemental densities into polylines to connect representative tissues, for which we took mass-weighted mean for the tissues in limited density regions. We compared the polyline functions of mass density with a bi-line for electron density and broken lines for elemental densities, which were derived from preceding studies. Results: There was generally high correlation between mass density and the other densities except of C, N, and O for light spongiosa tissues occupying 1% of body mass. The polylines fitted to the dominant tissues and were generally consistent with the bi-line and the broken lines. Conclusions: We have formulated the invariant relations between mass and electron densities and from mass to elemental densities for body tissues. The formulation enables Monte Carlo simulation in treatment planning practice without additional burden with CT-number calibration.