Photoemission System with Polarized Hard X-rays for Probing Ground State Symmetry of Strongly Correlated Materials


الملخص بالإنكليزية

We have developed a polarized hard X-ray photoemission (HAXPES) system to study the ground-state symmetry of strongly correlated materials. The linear polarization of the incoming X-ray beam is switched by the transmission-type phase retarder composed of two diamond (100) crystals. The best degree of the linear polarization $P_L$ is $-0.96$, containing the vertical polarization component of 98%. A newly developed low temperature two-axis manipulator enables easy polar and azimuthal rotations to select the detection direction of photoelectrons. The lowest temperature achieved is 9 K, offering us a chance to access the ground state even for the strongly correlated electron systems in cubic symmetry. The co-axial sample monitoring system with the long-working-distance microscope enables us to keep measuring the same region on the sample surface before and after rotation procedures. Combining this sample monitoring system with a micro-focused X-ray beam by means of an ellipsoidal Kirkpatrick-Baez mirror (25 $mu$m $times$ 25 $mu$m (FWHM)), we have demonstrated the polarized valence-band HAXPES on NiO for voltage application as resistive random access memories to reveal the origin of the metallic spectral weight near the Fermi level.