The method of extraction of the $e/h$ ratio, the degree of non-compensation, of the electromagnetic compartment of the combined calorimeter is suggested. The $e/h$ ratio of $1.74pm0.04$ has been determined on the basis of the 1996 combined calorimeter test beam data. This value agrees with the prediction that $e/h > 1.7$ for this electromagnetic calorimeter.
The e/$pi$ ratio for the Barrel Combined Calorimeter Prototype, composed from electromagnetic LAr calorimeter and hadronic Tile calorimter was investigated. Response of Combined Calorimeter on pions and electrons in the energy region 20 - 300 GeV was
The detailed experimental information about the electron and pion responses, the electron energy resolution and the e/h ratio as a function of incident energy E, impact point Z and incidence angle $Theta$ of the Module-0 of the iron-scintillator barr
This work is devoted to the experimental study of the longitudinal hadronic shower development in the ATLAS barrel combined prototype calorimeter consisting of the lead-liquid argon electromagnetic part and the iron-scintillator hadronic part. The re
The new simple method of the energy reconstruction for a combined calorimeter, which we called the e/h method, is suggested. It uses only the known e/h ratios and the electron calibration constants and does not require the determination of any parame
The method of extraction of the e/h ratio for an electromagnetic compartment of a combined calorimeter is suggested and the non-compensation was determined. The results agree with the Monte Carlo prediction and results of the weighting method for ele