ترغب بنشر مسار تعليمي؟ اضغط هنا

Direct positron emission imaging: ultra-fast timing enables reconstruction-free imaging

157   0   0.0 ( 0 )
 نشر من قبل Simon Cherry
 تاريخ النشر 2021
  مجال البحث فيزياء
والبحث باللغة English
 تأليف Ryosuke Ota




اسأل ChatGPT حول البحث

Positron emission tomography, like many other tomographic imaging modalities, relies on an image reconstruction step to produce cross-sectional images from projection data. Detection and localization of the back-to-back annihilation photons produced by positron-electron annihilation defines the trajectories of these photons, which when combined with tomographic reconstruction algorithms, permits recovery of the distribution of positron-emitting radionuclides. Here we produce cross-sectional images directly from the detected coincident annihilation photons, without using a reconstruction algorithm. Ultra-fast radiation detectors with a resolving time averaging 32 picoseconds measured the difference in arrival time of pairs of annihilation photons, localizing the annihilation site to 4.8 mm. This is sufficient to directly generate an image without reconstruction and without the geometric and sampling constraints that normally present for tomographic imaging systems.



قيم البحث

اقرأ أيضاً

T2-Shuffling reconstructs multiple sharp T2-weighted images from a single volumetric fast spin-echo (3D-FSE) scan. Wave-CAIPI is a parallel imaging technique that achieves good reconstruction at high accelerations through additional sinusoidal gradie nts that induce a voxel spreading effect in the readout direction to better take advantage of coil-sensitivity information. In this work, the Shuffling model in T2-Shuffling is augmented with wave-encoding to achieve higher acceleration capability. The resulting Wave-Shuffling approach is applied to 3D-FSE and Magnetization-Prepared Rapid Gradient-Echo (MPRAGE) to achieve rapid, 1 mm-isotropic resolution, time-resolved structural imaging.
Optical tomographic imaging of biological specimen bases its reliability on the combination of both accurate experimental measures and advanced computational techniques. In general, due to high scattering and absorption in most of the tissues, multi view geometries are required to reduce diffuse halo and blurring in the reconstructions. Scanning processes are used to acquire the data but they inevitably introduces perturbation, negating the assumption of aligned measures. Here we propose an innovative, registration free, imaging protocol implemented to image a human tumor spheroid at mesoscopic regime. The technique relies on the calculation of autocorrelation sinogram and object autocorrelation, finalizing the tomographic reconstruction via a three dimensional Gerchberg Saxton algorithm that retrieves the missing phase information. Our method is conceptually simple and focuses on single image acquisition, regardless of the specimen position in the camera plane. We demonstrate increased deep resolution abilities, not achievable with the current approaches, rendering the data alignment process obsolete.
213 - C. Lang , D. Habs , K. Parodi 2013
We present a nuclear medical imaging technique, employing triple-gamma trajectory intersections from beta^+ - gamma coincidences, able to reach sub-millimeter spatial resolution in 3 dimensions with a reduced requirement of reconstructed intersection s per voxel compared to a conventional PET reconstruction analysis. This $gamma$-PET technique draws on specific beta^+ - decaying isotopes, simultaneously emitting an additional photon. Exploiting the triple coincidence between the positron annihilation and the third photon, it is possible to separate the reconstructed true events from background. In order to characterize this technique, Monte-Carlo simulations and image reconstructions have been performed. The achievable spatial resolution has been found to reach ca. 0.4 mm (FWHM) in each direction for the visualization of a 22Na point source. Only 40 intersections are sufficient for a reliable sub-millimeter image reconstruction of a point source embedded in a scattering volume of water inside a voxel volume of about 1 mm^3 (high-resolution mode). Moreover, starting with an injected activity of 400 MBq for ^76Br, the same number of only about 40 reconstructed intersections are needed in case of a larger voxel volume of 2 x 2 x 3~mm^3 (high-sensitivity mode). Requiring such a low number of reconstructed events significantly reduces the required acquisition time for image reconstruction (in the above case to about 140 s) and thus may open up the perspective for a quasi real-time imaging.
A detector using liquid Xenon (LXe) in the scintillation mode is studied for Positron Emission Tomography (PET) of small animals. Its specific design aims at taking full advantage of the Liquid Xenon scintillation properties. This paper reports on en ergy, time and spatial resolution capabilities of the first LXe prototype module equipped with a Position Sensitive Photo- Multiplier tube (PSPMT) operating in the VUV range (178 nm) and at 165 K. The experimental results show that such a LXe PET configuration might be a promising solution insensitive to any parallax effect.
82 - Xin Shu , Sameera Sansre , Di Jin 2020
Leukocyte differential test is a widely performed clinical procedure for screening infectious diseases. Existing hematology analyzers require labor-intensive work and a panel of expensive reagents. Here we report an artificial-intelligence enabled re agent-free imaging hematology analyzer (AIRFIHA) modality that can accurately classify subpopulations of leukocytes with minimal sample preparation. AIRFIHA is realized through training a two-step residual neural network using label-free images of separated leukocytes acquired from a custom-built quantitative phase microscope. We validated the performance of AIRFIHA in randomly selected test set and cross-validated it across all blood donors. AIRFIHA outperforms current methods in classification accuracy, especially in B and T lymphocytes, while preserving the natural state of cells. It also shows a promising potential in differentiating CD4 and CD8 cells. Owing to its easy operation, low cost, and strong discerning capability of complex leukocyte subpopulations, we envision AIRFIHA is clinically translatable and can also be deployed in resource-limited settings, e.g., during pandemic situations for the rapid screening of infectious diseases.