The response of a position-sensitive Li-glass scintillator detector being developed for thermal-neutron detection with 6 mm position resolution has been investigated using collimated beams of thermal neutrons. The detector was moved perpendicularly through the neutron beams in 0.5 to 1.0 mm horizontal and vertical steps. Scintillation was detected in an 8 X 8 pixel multi-anode photomultiplier tube on an event-by-event basis. In general, several pixels registered large signals at each neutron-beam location. The number of pixels registering signal above a set threshold was investigated, with the maximization of the single-hit efficiency over the largest possible area of the detector as the primary goal. At a threshold of ~50% of the mean of the full-deposition peak, ~80% of the events were registered in a single pixel, resulting in an effective position resolution of ~5 mm in X and Y. Lower thresholds generally resulted in events demonstrating higher pixel multiplicities, but these events could also be localized with ~5 mm position resolution.
The response of a position-sensitive Li-glass based scintillation detector to focused beams of 2.5 MeV protons and deuterons has been investigated. The beams were scanned across the detector in 0.5 mm horizontal and vertical steps perpendicular to th
The response of a position-sensitive Li-glass scintillator detector to $alpha$-particles from a collimated $^{241}$Am source scanned across the face of the detector has been measured. Scintillation light was read out by an 8 X 8 pixel multi-anode pho
We have studied neutron response of PARIS phoswich [LaBr$_3$(Ce)-NaI(Tl)] detector which is being developed for measuring the high energy (E$_{gamma}$ = 5 - 30 MeV) $gamma$ rays emitted from the decay of highly collective states in atomic nuclei. The
In the Kaos spectrometer at the Mainz Microtron a high-resolution coordinate detector for high-energy particles is operated. It consists of scintillating fibres with diameters of < 1mm and is read out by > 4000 multi-anode photomultiplier channels. I
The aim of this paper is to fully characterize the new multi-anode photomultiplier tube R11265-103-M64, produced by Hamamatsu. Its high effective active area (77%), its pixel size, the low dark signal rate and the capability to detect single photon s