Design and Performance of the X-ray polarimeter X-Calibur

326   0   0.0 ( 0 )
 Added by Matthias Beilicke
 Publication date 2014
  fields Physics
and research's language is English




Ask ChatGPT about the research

X-ray polarimetry promises to give qualitatively new information about high-energy astrophysical sources, such as binary black hole systems, micro-quasars, active galactic nuclei, neutron stars, and gamma-ray bursts. We designed, built and tested a X-ray polarimeter, X-Calibur, to be used in the focal plane of the balloon-borne InFOCuS grazing incidence X-ray telescope. X-Calibur combines a low-Z scatterer with a CZT detector assembly to measure the polarization of 20-80keV X-rays making use of the fact that polarized photons scatter preferentially perpendicular to the electric field orientation. X-Calibur achieves a high detection efficiency of ~80%. The X-Calibur detector assembly is completed, tested, and fully calibrated. The response to a polarized X-ray beam was measured successfully at the Cornell High Energy Synchrotron Source. This paper describes the design, calibration and performance of the X-Calibur polarimeter. In principle, a similar space-borne scattering polarimeter could operate over the broader 2-100keV energy band.



rate research

Read More

X-ray polarimetry promises to give qualitatively new information about high-energy astrophysical sources, such as binary black hole systems, micro-quasars, active galactic nuclei, and gamma-ray bursts. We designed, built and tested a hard X-ray polarimeter X-Calibur to be used in the focal plane of the InFOCuS grazing incidence hard X-ray telescope. X-Calibur combines a low-Z Compton scatterer with a CZT detector assembly to measure the polarization of 10-80 keV X-rays making use of the fact that polarized photons Compton scatter preferentially perpendicular to the electric field orientation. X-Calibur achieves a high detection efficiency of order unity.
X-ray polarimetry promises to give new information about high-energy astrophysical sources, such as binary black hole systems, micro-quasars, active galactic nuclei, and gamma-ray bursts. We designed, built and tested a hard X-ray polarimeter X-Calibur to be used in the focal plane of the InFOCuS grazing incidence hard X-ray telescope. X-Calibur combines a low-Z Compton scatterer with a CZT detector assembly to measure the polarization of 10-80 keV X-rays making use of the fact that polarized photons Compton scatter preferentially perpendicular to the electric field orientation. X-Calibur achieves a high detection efficiency of order unity.
X-ray polarimetry has seen a growing interest in recent years. Improvements in detector technology and focusing X-ray optics now enable sensitive astrophysical X-ray polarization measurements. These measurements will provide new insights into the processes at work in accreting black holes, the emission of X-rays from neutron stars and magnetars, and the structure of AGN jets. X-Calibur is a balloon-borne hard X-ray scattering polarimeter. An X-ray mirror with a focal length of 8 m focuses X-rays onto the detector, which consists of a plastic scattering element surrounded by Cadmium-Zinc-Telluride detectors, which absorb and record the scattered X-rays. Since X-rays preferentially scatter perpendicular to their polarization direction, the polarization properties of an X-ray beam can be inferred from the azimuthal distribution of scattered X-rays. A close alignment of the X-ray focal spot with the center of the detector is required in order to reduce systematic uncertainties and to maintain a high photon detection efficiency. This places stringent requirements on the mechanical and thermal stability of the telescope structure. During the flight on a stratospheric balloon, X-Calibur makes use of the Wallops Arc-Second Pointer (WASP) to point the telescope at astrophysical sources. In this paper, we describe the design, construction, and test of the telescope structure, as well as its performance during a 25-hour flight from Ft. Sumner, New Mexico. The carbon fiber-aluminum composite structure met the requirements set by X-Calibur and its design can easily be adapted for other types of experiments, such as X-ray imaging or spectroscopic telescopes.
The performance of the Time Projection Chamber (TPC) polarimeter for the Polarimeter for Relativistic Astrophysical X-ray Sources (PRAXyS) Small Explorer was evaluated using polarized and unpolarized X-ray sources. The PRAXyS mission will enable exploration of the universe through X-ray polarimetry in the 2-10 keV energy band. We carried out performance tests of the polarimeter at the Brookhaven National Laboratory, National Synchrotron Light Source (BNL-NSLS) and at NASAs Goddard Space Flight Center. The polarimeter was tested with linearly polarized, monochromatic X-rays at 11 different energies between 2.5 and 8.0 keV. At maximum sensitivity, the measured modulation factors at 2.7, 4.5 and 8.0 keV are 27%, 43% and 59%, respectively and the measured angle of polarization is consistent with the expected value at all energies. Measurements with a broadband, unpolarized X-ray source placed a limit of less than 1% on false polarization in the PRAXyS polarimeter.
In the 50 years since the advent of X-ray astronomy there have been many scientific advances due to the development of new experimental techniques for detecting and characterising X-rays. Observations of X-ray polarisation have, however, not undergone a similar development. This is a shortcoming since a plethora of open questions related to the nature of X-ray sources could be resolved through measurements of the linear polarisation of emitted X-rays. The PoGOLite Pathfinder is a balloon-borne hard X-ray polarimeter operating in the 25 - 240 keV energy band from a stabilised observation platform. Polarisation is determined using coincident energy deposits in a segmented array of plastic scintillators surrounded by a BGO anticoincidence system and a polyethylene neutron shield. The PoGOLite Pathfinder was launched from the SSC Esrange Space Centre in July 2013. A near-circumpolar flight was achieved with a duration of approximately two weeks. The flight performance of the Pathfinder design is discussed for the three Crab observations conducted. The signal-to-background ratio for the observations is shown to be 0.25$pm$0.03 and the Minimum Detectable Polarisation (99% C.L.) is (28.4$pm$2.2)%. A strategy for the continuation of the PoGOLite programme is outlined based on experience gained during the 2013 maiden flight.