Performance of a large TeO2 crystal as a cryogenic bolometer in searching for neutrinoless double beta decay


Abstract in English

Bolometers are ideal devices in the search for neutrinoless Double Beta Decay. Enlarging the mass of individual detectors would simplify the construction of a large experiment, but would also decrease the background per unit mass induced by alpha-emitters located close to the surfaces and background arising from external and internal gammas. We present the very promising results obtained with a 2.13 kg TeO2 crystal. This bolometer, cooled down to a temperature of 10.5 mK in a dilution refrigerator located deep underground in the Gran Sasso National Laboratories, represents the largest thermal detector ever operated. The detector exhibited an energy resolution spanning a range from 3.9 keV (at 145 keV) to 7.8 keV (at the 2615 gamma-line of 208Tl) FWHM. We discuss the decrease in the background per unit mass that can be achieved increasing the mass of a bolometer.

Download