Total Gas Pressure Dependence on operation time for Self-Quenching Geiger-Mueller Counters
published by Aِl-Baath University
in 2016
in
and research's language is
العربية
Download
Abstract in English
In this research, we have derived an analytical equation gives
dependency of the total gas pressure to operation time for halogen
self-quenched GM counter tubes and organic vapor or gas selfquenched
GM counter tubes. This equation indicates that the gas
pressure exponentially decreases as the operation time increases, at
constant temperature of tube and constant activity that this tube
exposed to.
References used
(S. A. Korff, W. Spatz, and N. Hilberry, Rev. Sci. inst. 13, 127 (1942
W. D. B. SPATZ. PHYSICAL REVIEW VOLUME 64, NUMBERS 7 AND 8 OCTOBER 1 AND 15, 1943
Herbert Fridman, Geiger Counter Tubes, PROCEEDINGS OF THE I.R.E.— Waves and Electrons Section, 791-808 July,1949