Calculation of dose distributions produced by 6 MeV photon beam at variable depth in 3D Water Phantom using the MCNP4C2


Abstract in English

To calculate the dose distributions of 6 MeV photon beam at variable depth in 3D water phantom the MCNP4C2 code was used, and the simulated dose profile was compared with that of the treatment planning computer system (TPS), and a good agreement was found between them. In conclusion, the MCNP4C2 code package presents a good tool adaptable to get dose distributions for the 6MeV photon beam and it can be considered as confirmed method for patient dose calculations.

References used

Papanikolaou N., Battista J., Boyer A., Kappas C., Klein E., Mackie T.R., Sharpe M. and Van Dyk J., 2004. Tissue Inhomogeneity Corrections for Megavoltage Photon Beams. AAPM Report No 85, Task Group No 65 of the Radiation Therapy Committee of the American Association of Physicist in Medicine. Madison: Medical Physics Publishing
Kawrakow, I., 2000. Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version. Med Phys., 27: 485- 498
Rogers, D.W., 2006. Fifty years of Monte Carlo simulations for medical physics. Phys Med Biol., 51: R287-301

Download