Determination of Cholesterol in some Food Stuff by High performance liquid chromatography Technique


Abstract in English

The purpose of this article is how to use HPLC technique for quantitive analysis of cholesterol in some foods. The best conditions are studied for this analysis including composition of mobile phase, temperature, flow rate, the wavelength of detector UV-VIS, and acidity of mobile phase. The column used for cholesterol determination by this method was: C18(5μm,250x4.6mm) with injection volume of 20μL. Methanol 100% was used as mobile phase.

References used

DAVID.H, 2000- Modern Analytical Chemistry, Journal of International Edition-North America
UFFE Ravnskov. M. D, 2000- The Cholesterol Myths, New Trends –Washington
DENNIS. E, VANCE. H, VAN. D.B, 2000- Journal of Biochimica et BiophysicaActa (BBA),1529 (1-3), 1-8