Study of the Effect of Electrostatic Charge on the Delivered Dose of Metered Dose Inhaler


Abstract in English

A study of the effect of electrostatic charge that is generated in actuator due to drying methods on the delivered dose of metered dose inhaler has been done in tow cases: Washing actuator and drying it by a clothe. Washing actuator and let it dry alone. A study has been done on metered dose inhaler containing salbutmaol sulfate and manufactured by a local company. The results show the importance of drying methods on the delivered dose, the delivered doses of metered dose inhaler after washing the actuator and dry it by a clothe hadn't been achieved the USP-34 requirements (out of range 75-125%), while the delivered doses after washing actuator and let it dry alone had been achieved the USP-34 requirements. The difference in results between the two cases belong to electrostatic charge generated in actuator due to drying methods

References used

Anhoj, J.;Bisgaard, H.;Lipworth.B.J. Effect of Electrostatic Charge in Plastic Spacers on the Lung Delivery of HFA-Salbutamol in Children. Br J Pharmacol.47(3),1999,333-336
Anouk, C.A.; Sennbauser, F.H.; Wildhaber, J.H. Factor Affecting the Efficiency of Aerosol Therapy with Pressurized Metered Dose Inhalers through Plastic Spacers. SwissMed Wlky,2001,14-18
SwissMed Wlky,2001,14-18. 3. Deveswaran, R.; Bharath, S.; Basavaraj, BV.; Abraham, S.; Furtado, SH.; Madhavan, V. Concept and Techniques of Pharmaceutical Powder Mixing Process: A Current Update. Research J.Pharm and Tech.2(2).April,2009,245-249
Kwok, P.C.; Chan, H.H. Electrostatic of Pharmaceutical Aerosols. J Pharma Pharmacol.Dec;61(12),2009,1587-99.Doi:10.1211/jpp/61.12.0002
Mitchell, J.P.; Coppolo, D.P.; Nagel, M.W. Electrostatics and Inhaled Medications: Influence on Delivery via Pressurized Metered Dose Inhalers and Add on Devices. Respir Care,2007,Mar;52(3),283-300