In Vitro Study aboutthe Effect of Dual-Thread and Diameter onthe Orthodontic Miniscrews Primary Stability


Abstract in English

The aim of the study was to determine the effect of Dual-Thread and diameter on the orthodontic miniscrew primary stability, and define the best shape and diameter of miniescrew can be used to give the best stability. The study sample consisted of (24) orthodontic miniscrews made of (Ti-6Al-4V) alloy with 2 shapes (regular and Dual-Thread) , and 2 diameters (1.6 , 1.8mm). The experiment has been divided into 3 tests : measuring the maximum insertion torque , the Periotest value , the maximum removal torque. The results showed that the maximum insertion and removal torque were statistically higher while the Periotest values were statistically lower when using the Dual-Thread and the diameter of 1.8 mm than the regular shape and the diameter of 1.6 mm(P>0.01)

References used

GRABER TM, VANARSDALL RL. Orthodontics – principles and techniques.second ed. Rio de Janeiro: Guanabara Koogan; 1996
FAVERO L, BROLLO P, BRESSAN E. Orthodontic anchorage with specific fixtures: related study analysis. Am J OrthodDentofacialOrthop 2002;122:84–94
MELSEN, B. and VERNA, C. Miniscrew Implants: The Aarhus Anchorage System. SeminOrthod, Vol 11, 2005, 24-31
NANDA, R. and URIBE, F.A. Temporary Anchorage Devices. Mosby Elsevier, China, 2009
ZHAO L., XU,B Z, YANG Z, WEI X, TANG T, ZHAO Z.Orthodontic mini-implant stability in different healing times before loading: A microscopic computerized tomographic and biomechanical analysis. Oral Surg Oral Med Oral Pathol Oral RadiolEndod 2009;108:196-202