Dead-time correction for spectroscopic photon counting pixel detectors


Abstract in English

Modern photon counting pixel detectors enabled a revolution in applications at synchrotron light sources and beyond in the last decade. One of the limitations of current detectors is reduced counting linearity or even paralysis at high counting rates, due to dead-time which results in photon pile-up. Existing dead-time and pile-up models fail to reproduce the complexity of dead-time effects on photon counting, resulting in empirical calibrations for particular detectors at best, imprecise linearization methods, or no linearization. This problem will increase in the future as many synchrotron light sources plan significant brilliance upgrades and free-electron lasers plan moving to a quasi-continuous operation mode. We present here the first models that use the actual behavior of the analog pre-amplifiers in spectroscopic photon counting pixel detectors with constant current discharge (e.g., Medipix family of detectors) to deduce more accurate analytical models and optimal linearization methods. In particular, for detectors with at least two counters per pixel, we completely eliminate the need of calibration, or previous knowledge of the detector and beam parameters (dead-time, integration time, large sets of synchrotron filling patterns). This is summarized in several models with increasing complexity and accuracy. Finally, we present a general empirical approach applicable to any particular cases where the analytical approach is not sufficiently precise.