Spatial resolution measurement of Triple-GEM detector and diffraction imaging test at synchrotron radiation


Abstract in English

A triple-GEM detector with two-dimensional readout is developed. The detector provides high position resolution for powder diffraction experiments at synchrotron radiation. Spatial resolution of the detector is measured in the lab using a 55Fe X-ray source. A resolution of about 110 um FWHM is achieved. The energy resolution is better than 27% for 5.9 keV X-rays. The detectors validity under illumination of photons in particular energy range is verified using a Cu X-ray tube. Imaging of the head of a wire stripper with X-ray tube demonstrates its imaging ability. A diffraction imaging experiment using the sample of powder SiO2 is successfully carried out at 1W2B laboratory of Beijing Synchrotron Radiation Facility (BSRF). Different diffraction rings are clearly seen under various X-ray energies.

Download