ﻻ يوجد ملخص باللغة العربية
We report on the successful operation of a double phase Liquid Argon Large Electron Multiplier Time Projection Chamber (LAr LEM-TPC) equipped with two dimensional projective anodes with dimensions 10$times$10 cm$^2$, and with a maximum drift length of 21 cm. The anodes were manufactured for the first time from a single multilayer printed circuit board (PCB). Various layouts of the readout views have been tested and optimised. In addition, the ionisation charge was efficiently extracted from the liquid to the gas phase with a single grid instead of two previously. We studied the response and the gain of the detector to cosmic muon tracks. To study long-term stability over several weeks, we continuously operated the chamber at fixed electric field settings. We reproducibly observe that after an initial decrease with a characteristic time of $tauapprox 1.6$ days, the observed gain is stable. In 46 days of operation, a total of 14.6 million triggers have been collected at a stable effective gain of $G_inftysim 15$ corresponding to a signal-to-noise ratio $(S/N)gtrsim 60$ for minimum ionising tracks. During the full period, eight discharges across the LEM were observed. A maximum effective gain of 90 was also observed, corresponding to a signal-to-noise ratio $(S/N)gtrsim 400$ for minimum ionising tracks, or $S/Napprox10$ for an energy deposition of 15 keV on a single readout channel.
We successfully operated a novel kind of LAr Time Projection Chamber based on a Large Electron Multiplier (LEM) readout system. The prototype, of about 3 liters active volume, is operated in liquid-vapour (double) phase pure Ar. The ionization electr
In this paper we describe the design, construction, and operation of a first large area double-phase liquid argon Large Electron Multiplier Time Projection Chamber (LAr LEM-TPC). The detector has a maximum drift length of 60 cm and the readout consis
The Large Electron Multipliers (LEMs) are key components of double phase liquid argon TPCs. The drifting charges after being extracted from the liquid are amplified in the LEM positioned half a centimeter above the liquid in pure argon vapor at 87 K.
In this paper we present results from a test of a small Liquid Argon Large Electron Multiplier Time Projection Chamber (LAr LEM-TPC). This detector concept provides a 3D-tracking and calorimetric device capable of charge amplification, suited for nex
Gated wires are widely used in Time Projection Chamber (TPC) to avoid ion back-flow (IBF) in the drift volume. The anode wires can provide stable gain at high voltage with a long lifetime. However, switching on and off the gated grid (GG) leads to a