ترغب بنشر مسار تعليمي؟ اضغط هنا

Timing Performance of a Micro-Channel-Plate Photomultiplier Tube

372   0   0.0 ( 0 )
 نشر من قبل Lukas Sohl
 تاريخ النشر 2019
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

The spatial dependence of the timing performance of the R3809U-50 Micro-Channel-Plate PMT (MCP-PMT) by Hamamatsu was studied in high energy muon beams. Particle position information is provided by a GEM tracker telescope, while timing is measured relative to a second MCP-PMT, identical in construction. In the inner part of the circular active area (radius r$<$5.5,mm) the time resolution of the two MCP-PMTs combined is better than 10~ps. The signal amplitude decreases in the outer region due to less light reaching the photocathode, resulting in a worse time resolution. The observed radial dependence is in quantitative agreement with a dedicated simulation. With this characterization, the suitability of MCP-PMTs as $text{t}_text{0}$ reference detectors has been validated.



قيم البحث

اقرأ أيضاً

We have developed a position response calibration method for a micro-channel plate (MCP) detector with a delay-line anode position readout scheme. Using an {em in situ} calibration mask, an accuracy of 8~$mu$m and a resolution of 85~$mu$m (FWHM) have been achieved for MeV-scale $alpha$ particles and ions with energies of $sim$10~keV. At this level of accuracy, the difference between the MCP position responses to high-energy $alpha$ particles and low-energy ions is significant. The improved performance of the MCP detector can find applications in many fields of AMO and nuclear physics. In our case, it helps reducing systematic uncertainties in a high-precision nuclear $beta$-decay experiment.
We studied the single-photoelectron detection capabilities of a multianode photomultiplier tube H8500C-03 and its performance in high magnetic field. Our results show that the device can readily resolve signals at the single photoelectron level makin g it suitable for photon detection in both threshold and ring imaging Cerenkov detectors. We also found that a large longitudinal magnetic field, up to 300 Gauss, induces a change in the relative output of at most 55% for an edge pixel, and of at most 15% for a central pixel. The H8500C-03 signal loss in transverse magnetic fields it is significantly more pronounced than for the longitudinal case. Our studies of single photoelectron reduction in magnetic fields point to the field induced misfocusing of the photoelectron extracted from the photocathode as primary cause of signal loss. With appropriate shielding this PMT could function in high magnetic field environments.
A new hybrid experiment has been constructed to measure the chemical composition of cosmic rays around the knee in the wide energy range by the Tibet AS$gamma$ collaboration at Tibet, China, since 2014. They consist of a high-energy air-shower-core a rray (YAC-II), a high-density air-shower array (Tibet-III) and a large underground water-Cherenkov muon-detector array (MD). In order to obtain the primary proton, helium and iron spectra and their knee positions in the energy range lower than $10^{16}$ eV, each of PMTs equipped to the MD cell is required to measure the number of photons capable of covering a wide dynamic range of 100 - $10^{6}$ photoelectrons (PEs) according to Monte Carlo simulations. In this paper, we firstly compare the characteristic features between R5912-PMT made by Japan Hamamatsu and CR365-PMT made by Beijing Hamamatsu. This is the first comparison between R5912-PMT and CR365-PMT. If there exists no serious difference, we will then add two 8-inch-in-diameter PMTs to meet our requirements in each MD cell, which are responsible for the range of 100 - 10000 PEs and 2000 - 1000000 PEs, respectively. That is, MD cell is expected to be able to measure the number of muons over 6 orders of magnitude.
129 - J. Brack , B. Delgado , J. Dhooghe 2012
Future large water Cherenkov and scintillator detectors have been proposed for measurements of long baseline neutrino oscillations, proton decay, supernova and solar neutrinos. To ensure cost-effectiveness and optimize scientific reach, one of the cr itical requirements for such detectors are large-area, high performance photomultiplier tubes (PMTs). One candidate for such a device is the Hamamatsu R11780, a 12 PMT that is available in both standard and high quantum efficien
207 - K.J. Ma , W.G. Kang , J.K. Ahn 2009
We have studied the afterpulse of a hemispherical photomultiplier tube for an upcoming reactor neutrino experiment. The timing, the amplitude, and the rate of the afterpulse for a 10 inch photomultiplier tube were measured with a 400 MHz FADC up to 1 6 ms time window after the initial signal generated by an LED light pulse. The time and amplitude correlation of the afterpulse shows several distinctive groups. We describe the dependencies of the afterpulse on the applied high voltage and the amplitude of the main light pulse. The present data could shed light upon the general mechanism of the afterpulse.
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا