A 20-Liter Test Stand with Gas Purification for Liquid Argon Research

463   0   0.0 ( 0 )
 نشر من قبل Yichen Li
 تاريخ النشر 2016
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

We describe the design of a 20-liter test stand constructed to study fundamental properties of liquid argon (LAr). This system utilizes a simple, cost-effective gas argon (GAr) purification to achieve high purity, which is necessary to study electron transport properties in LAr. An electron drift stack with up to 25 cm length is constructed to study electron drift, diffusion, and attachment at various electric fields. A gold photocathode and a pulsed laser are used as a bright electron source. The operational performance of this system is reported.



قيم البحث

اقرأ أيضاً

A filter system for removing electronegative impurities from liquid argon is described. The active components of the filter are adsorbing molecular sieve and activated-copper-coated alumina granules. The system is capable of purifying liquid argon to an oxygen-equivalent impurity concentration of better than 30 parts per trillion, corresponding to an electron drift lifetime of at least 10 ms. Reduction reactions that occur at about 250 degrees Celsius allow the filter material to be regenerated in-situ through a simple procedure. In the following work we describe the filter design, performance, and regeneration process.
The ARAPUCA is a novel concept for liquid argon scintillation light detection which has been proposed for the photon detection system of the Deep Underground Neutrino Experiment. The test in liquid argon of one of the first ARAPUCA prototypes is pres ented in this work, where the working principle is experimentally demonstrated. The prototype has an acceptance window of 9 cm$^2$ and is read-out by a single SiPM with active area of 0.36 cm$^2$. Its global detection efficiency was estimated by exposing it to a $^{238}U$ $alpha$ source and to cosmic rays and was found to be 1.15% $pm$ 0.15%, in good agreement with the prediction of a detailed Monte Carlo simulation of the device. Several other ARAPUCA prototypes of bigger dimensions and read-out by arrays of SiPMs have been built and are actually under test. In particular 32 ARAPUCA cells have been installed inside the protoDUNE detector, which is being assembled at CERN and will be operated in the second half of 2018.
We have built a gas-phase argon ionization detector to measure small nuclear recoil energies (< 10 keVee). In this paper, we describe the detector response to X-ray and gamma calibration sources, including analysis of pulse shapes, software triggers, optimization of gas content, and energy- and position-dependence of the signal. We compare our experimental results against simulation using a 5.9-keV X-ray source, as well as higher-energy gamma sources up to 1332 keV. We conclude with a description of the detector, DAQ, and software settings optimized for a measurement of the low-energy nuclear quenching factor in gaseous argon. This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory in part under Contract W-7405-Eng-48 and in part under Contract DE-AC52-07NA27344. Funded by Lab-wide LDRD. LLNL-JRNL-415990-DRAFT.
178 - Maf Alidra 2020
The rate of muons from LHC $pp$ collisions reaching the surface above the ATLAS interaction point is measured and compared with expected rates from decays of $W$ and $Z$ bosons and $b$- and $c$-quark jets. In addition, data collected during periods w ithout beams circulating in the LHC provide a measurement of the background from cosmic ray inelastic backscattering that is compared to simulation predictions. Data were recorded during 2018 in a 2.5 $times$ 2.5 $times$ 6.5~$rm{m}^3$ active volume MATHUSLA test stand detector unit consisting of two scintillator planes, one at the top and one at the bottom, which defined the trigger, and six layers of RPCs between them, grouped into three $(x,y)$-measuring layers separated by 1.74 m from each other. Triggers selecting both upward-going tracks and downward-going tracks were used.
218 - K. Mavrokoridis 2011
Future giant liquid argon (LAr) time projection chambers (TPCs) require a purity of better than 0.1 parts per billion (ppb) to allow the ionised electrons to drift without significant capture by any electronegative impurities. We present a comprehens ive study of the effects of electronegative impurity on gaseous and liquid argon scintillation light, an analysis of the efficacy of various purification chemicals, as well as the Liverpool LAr setup, which utilises a novel re-circulation purification system. Of the impurities tested - Air, O_2, H_2O, N_2 and CO_2 in the range of between 0.01 ppm to 1000 ppm - H_2O was found to have the most profound effect on gaseous argon scintillation light, and N_2 was found to have the least. Additionally, a correlation between the slow component decay time and the total energy deposited with 0.01 ppm - 100 ppm O_2 contamination levels in liquid argon has been established. The superiority of molecular sieves over anhydrous complexes at absorbing Ar gas, N_2 gas and H_2O vapour has been quantified using BET isotherm analysis. The efficiency of Cu and P_2O5 at removing O_2 and H_2O impurities from 1 bar N6 argon gas at both room temperature and -130 ^oC was investigated and found to be high. A novel, highly scalable LAr re-circulation system has been developed. The complete system, consisting of a motorised bellows pump operating in liquid and a purification cartridge, were designed and built in-house. The system was operated successfully over many days and achieved a re-circulation rate of 27 litres/hour and high purity.
التعليقات
جاري جلب التعليقات جاري جلب التعليقات
سجل دخول لتتمكن من متابعة معايير البحث التي قمت باختيارها
mircosoft-partner

هل ترغب بارسال اشعارات عن اخر التحديثات في شمرا-اكاديميا