ترغب بنشر مسار تعليمي؟ اضغط هنا

NIKA 2: next-generation continuum/polarized camera at the IRAM 30 m telescope and its prototype

65   0   0.0 ( 0 )
 نشر من قبل Alessia Ritacco
 تاريخ النشر 2016
  مجال البحث فيزياء
والبحث باللغة English




اسأل ChatGPT حول البحث

NIKA 2 (New Instrument of Kids Array) is a next generation continuum and polarized instrument successfully installed in October 2015 at the IRAM 30 m telescope on Pico-Veleta (Granada, Spain). NIKA 2 is a high resolution dual-band camera, operating with frequency multiplexed LEKIDs (Lumped Element Kinetic Inductance Detectors) cooled at 100 mK. Dual color images are obtained thanks to the simultaneous readout of a 1020 pixels array at 2 mm and 1140 x 2 pixels arrays at 1.15 mm with a final resolution of 18 and 12 arcsec respectively, and 6.5 arcmin of Field of View (FoV). The two arrays at 1.15 mm allow us to measure the linear polarization of the incoming light. This will place NIKA 2 as an instrument of choice to study the role of magnetic fields in the star formation process. The NIKA experiment, a prototype for NIKA 2 with a reduced number of detectors (about 400 LEKIDs) and FoV (1.8 arcmin), has been successfully operated at the IRAM 30 telescope in several open observational campaigns. The performance of the NIKA 2 polarization setup has been successfully validated with the NIKA prototype.



قيم البحث

اقرأ أيضاً

The New IRAM KID Array (NIKA) instrument is a dual-band imaging camera operating with Kinetic Inductance Detectors (KID) cooled at 100 mK. NIKA is designed to observe the sky at wavelengths of 1.25 and 2.14 mm from the IRAM 30 m telescope at Pico Vel eta with an estimated resolution of 13,arcsec and 18 arcsec, respectively. This work presents the performance of the NIKA camera prior to its opening to the astrophysical community as an IRAM common-user facility in early 2014. NIKA is a test bench for the final NIKA2 instrument to be installed at the end of 2015. The last NIKA observation campaigns on November 2012 and June 2013 have been used to evaluate this performance and to improve the control of systematic effects. We discuss here the dynamical tuning of the readout electronics to optimize the KID working point with respect to background changes and the new technique of atmospheric absorption correction. These modifications significantly improve the overall linearity, sensitivity, and absolute calibration performance of NIKA. This is proved on observations of point-like sources for which we obtain a best sensitivity (averaged over all valid detectors) of 40 and 14 mJy.s$^{1/2}$ for optimal weather conditions for the 1.25 and 2.14 mm arrays, respectively. NIKA observations of well known extended sources (DR21 complex and the Horsehead nebula) are presented. This performance makes the NIKA camera a competitive astrophysical instrument.
Millimeter-wave continuum astronomy is today an indispensable tool for both general Astrophysics studies and Cosmology. General purpose, large field-of-view instruments are needed to map the sky at intermediate angular scales not accessible by the hi gh-resolution interferometers and by the coarse angular resolution space-borne or ground-based surveys. These instruments have to be installed at the focal plane of the largest single-dish telescopes. In this context, we have constructed and deployed a multi-thousands pixels dual-band (150 and 260 GHz, respectively 2mm and 1.15mm wavelengths) camera to image an instantaneous field-of-view of 6.5arc-min and configurable to map the linear polarization at 260GHz. We are providing a detailed description of this instrument, named NIKA2 (New IRAM KID Arrays 2), in particular focusing on the cryogenics, the optics, the focal plane arrays based on Kinetic Inductance Detectors (KID) and the readout electronics. We are presenting the performance measured on the sky during the commissioning runs that took place between October 2015 and April 2017 at the 30-meter IRAM (Institut of Millimetric Radio Astronomy) telescope at Pico Veleta. NIKA2 has been successfully deployed and commissioned, performing in-line with the ambitious expectations. In particular, NIKA2 exhibits FWHM angular resolutions of around 11 and 17.5 arc-seconds at respectively 260 and 150GHz. The NEFD (Noise Equivalent Flux Densities) demonstrated on the maps are, at these two respective frequencies, 33 and 8 mJy*sqrt(s). A first successful science verification run has been achieved in April 2017. The instrument is currently offered to the astronomical community during the coming winter and will remain available for at least the next ten years.
149 - R. Adam , A. Adane , P. Ade 2014
High resolution observations of the thermal Sunyaev-Zeldovich (tSZ) effect are necessary to allow the use of clusters of galaxies as a probe for large scale structures at high redshifts. With its high resolution and dual-band capability at millimeter wavelengths, the NIKA camera can play a significant role in this context. NIKA is based on newly developed Kinetic Inductance Detectors (KIDs) and operates at the IRAM 30m telescope, Pico Veleta, Spain. In this paper, we give the status of the NIKA camera, focussing on the KID technology. We then present observations of three galaxy clusters: RX J1347.5-1145 as a demonstrator of the NIKA capabilities and the recent observations of CL J1226.9+3332 (z = 0.89) and MACS J0717.5+3745 (z = 0.55). We also discuss prospects for the final NIKA2 camera, which will have a 6.5 arcminute field of view with about 5000 detectors in two bands at 150 and 260 GHz.
The New IRAM KID Arrays 2 (NIKA2) consortium has just finished installing and commissioning a millimetre camera on the IRAM 30 m telescope. It is a dual-band camera operating with three frequency multiplexed kilo-pixels arrays of Lumped Element Kinet ic Inductance Detectors (LEKID) cooled at 150 mK, designed to observe the intensity and polarisation of the sky at 260 and 150 GHz (1.15 and 2 mm). NIKA2 is today an IRAM resident instrument for millimetre astronomy, such as Intra Cluster Medium from intermediate to distant clusters and so for the follow-up of Planck satellite detected clusters, high redshift sources and quasars, early stages of star formation and nearby galaxies emission. We present an overview of the instrument performance as it has been evaluated at the end of the commissioning phase.
NIKA2 is a dual-band millimetric continuum camera of 2900 Kinetic Inductance Detectors (KID), operating at $150$ and $260,rm{GHz}$, installed at the IRAM 30-meter telescope. We present the performance assessment of NIKA2 after one year of observation using a dedicated point-source calibration method, referred to as the emph{baseline} method. Using a large data set acquired between January 2017 and February 2018 that span the whole range of observing elevations and atmospheric conditions encountered at the IRAM 30-m telescope, we test the stability of the performance parameters. We report an instantaneous field of view (FOV) of 6.5 in diameter, filled with an average fraction of $84%$ and $90%$ of valid detectors at $150$ and $260,rm{GHz}$, respectively. The beam pattern is characterized by a FWHM of $17.6 pm 0.1$ and $11.1pm 0.2$, and a beam efficiency of $77% pm 2%$ and $55% pm 3%$ at $150$ and $260,rm{GHz}$, respectively. The rms calibration uncertainties are about $3%$ at $150,rm{GHz}$ and $6%$ at $260,rm{GHz}$. The absolute calibration uncertainties are of $5%$ and the systematic calibration uncertainties evaluated at the IRAM 30-m reference Winter observing conditions are below $1%$ in both channels. The noise equivalent flux density (NEFD) at $150$ and $260,rm{GHz}$ are of $9 pm 1, rm{mJy}cdot s^{1/2}$ and $30 pm 3, rm{mJy}cdot s^{1/2}$. This state-of-the-art performance confers NIKA2 with mapping speeds of $1388 pm 174$ and $111 pm 11 ,rm{arcmin}^2cdot rm{mJy}^{-2}cdot rm{h}^{-1}$ at $150$ and $260,rm{GHz}$. With these unique capabilities of fast dual-band mapping at high (better that 18) angular resolution, NIKA2 is providing an unprecedented view of the millimetre Universe.