Laboratory Studies on the Removal of Radon-Born Lead from KamLANDs Organic Liquid Scintillator


الملخص بالإنكليزية

The removal of radioactivity from liquid scintillator has been studied in preparation of a low background phase of KamLAND. This paper describes the methods and techniques developed to measure and efficiently extract radon decay products from liquid scintillator. We report the radio-isotope reduction factors obtained when applying various extraction methods. During this study, distillation was identified as the most efficient method for removing radon daughters from liquid scintillator.

تحميل البحث